Alignment of a volumetric tomography system

被引:22
作者
Stevens, GM [1 ]
Saunders, R
Pelc, NJ
机构
[1] Stanford Univ, Dept Radiol, Stanford, CA 94305 USA
[2] Stanford Univ, Dept Appl Phys, Stanford, CA 94305 USA
[3] Gen Elect Med Syst, Waukesha, WI 53188 USA
关键词
volumetric tomography; tomosynthesis; system alignment; interpolation;
D O I
10.1118/1.1382609
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A test-bed system has been developed for imaging phantoms with tomosynthesis and volumetric computed tomography. This system incorporates an amorphous silicon flat panel detector on a movable gantry and a computer-controlled rotational positioning stage. In this paper, an analysis of the sensitivity of reconstructed images to geometrical misalignment is presented. Application of this method to circular digital tomosynthesis is examined, with spatial resolution in the focal plane as the criterion for evaluating the effect of misalignment. A software-based method is presented for correcting data for imperfect system alignment prior to image reconstruction. Experimental results yield reconstructed images with spatial resolution approaching the theoretical limit based on detector pixel size and accounting for data interpolation. (C) 2001 American Association of Physicists in Medicine.
引用
收藏
页码:1472 / 1481
页数:10
相关论文
共 22 条
[1]   Empirical investigation of the signal performance of a high-resolution, indirect detection, active matrix flat-panel imager (AMFPI) for fluoroscopic and radiographic operation [J].
Antonuk, LE ;
ElMohri, Y ;
Siewerdsen, JH ;
Yorkston, J ;
Huang, W ;
Scarpine, VE ;
Street, RA .
MEDICAL PHYSICS, 1997, 24 (01) :51-70
[2]   Virtual alignment of x-ray cone-beam tomography system using two calibration aperture measurements [J].
Bronnikov, AV .
OPTICAL ENGINEERING, 1999, 38 (02) :381-386
[3]   CT FAN BEAM RECONSTRUCTION WITH A NONSTATIONARY AXIS OF ROTATION [J].
CONCEPCION, JA ;
CARPINELLI, JD ;
KUOPETRAVIC, G ;
REISMAN, S .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (01) :111-116
[4]   RECONSTRUCTION FOR FAN BEAM WITH AN ANGULAR-DEPENDENT DISPLACED CENTER-OF-ROTATION [J].
CRAWFORD, CR ;
GULLBERG, GT ;
TSUI, BMW .
MEDICAL PHYSICS, 1988, 15 (01) :67-71
[5]  
CURRY TS, 1981, INTRO PHYSICS DIAGNO
[6]  
DOBBINS JT, 1998, RADIOLOGY 1998 SCI S, P280
[7]   Evaluation of imaging geometries calculated from biplane images [J].
Esthappan, J ;
Harauchi, H ;
Hoffmann, KR .
MEDICAL PHYSICS, 1998, 25 (06) :965-975
[8]   Three-dimensional computed tomographic reconstruction using a C-arm mounted XRII: Image-based correction of gantry motion nonidealities [J].
Fahrig, R ;
Holdsworth, DW .
MEDICAL PHYSICS, 2000, 27 (01) :30-38
[9]  
FAHRIG R, 1996, P SOC PHOTO-OPT INS, V2708, P30
[10]   Digital mammography [J].
Feig, SA ;
Yaffe, MJ .
RADIOGRAPHICS, 1998, 18 (04) :893-901