High-resolution dosimetry using radiochromic film and a document scanner

被引:51
作者
Stevens, MA [1 ]
Turner, JR [1 ]
Hugtenburg, RP [1 ]
Butler, PH [1 ]
机构
[1] CHRISTCHURCH HOSP, DEPT MED PHYS & BIOENGN, CHRISTCHURCH, NEW ZEALAND
关键词
D O I
10.1088/0031-9155/41/11/008
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
A method of reading exposed radiochromic film is described which has significant advantages over conventional densitometry. The method employs a document scanner and associated software for imaging the film. The resulting images are easily analysed using standard software to yield high-resolution dose maps. A calibration was performed which relates scanner signal to dose, allowing for the determination of dose at any point on an exposed film. Results obtained using a broad-band densitometer are compared to those where the scanner has been used. The technique was used to measure the dose distribution around a COMS-type ophthalmic applicator.
引用
收藏
页码:2357 / 2365
页数:9
相关论文
共 16 条
[1]   MONTE-CARLO SIMULATION OF THE DOSE DISTRIBUTION AROUND I-125 SEEDS [J].
BURNS, GS ;
RAESIDE, DE .
MEDICAL PHYSICS, 1987, 14 (03) :420-424
[2]   HIGH-SENSITIVITY GAFCHROMIC FILM DOSIMETRY FOR I-125 SEED [J].
CHIUTSAO, ST ;
DELAZERDA, A ;
LIN, J ;
KIM, JH .
MEDICAL PHYSICS, 1994, 21 (05) :651-657
[3]  
CHU RDH, 1990, RADIAT PHYS CHEM, V35, P767
[4]   I-125 eye plaque dose distribution including penumbra characteristics [J].
delaZerda, A ;
ChiuTsao, ST ;
Lin, J ;
Boulay, LL ;
Kanna, I ;
Kim, JH ;
Tsao, HS .
MEDICAL PHYSICS, 1996, 23 (03) :407-418
[5]   BACKSCATTER FACTORS FOR X-RAYS GENERATED AT VOLTAGES BETWEEN 10 AND 100 KV [J].
GROSSWENDT, B .
PHYSICS IN MEDICINE AND BIOLOGY, 1984, 29 (05) :579-591
[6]   THE 3-DIMENSIONAL DOSE DISTRIBUTION OF I-125 SEEDS IN TISSUE [J].
HARTMANN, GH ;
SCHLEGEL, W ;
SCHARFENBERG, H .
PHYSICS IN MEDICINE AND BIOLOGY, 1983, 28 (06) :693-699
[7]   MONTE-CARLO CALCULATION OF ENERGY BUILDUP FACTORS IN THE RANGE FROM 15KEV TO 100KEV, WITH SPECIAL REFERENCE TO THE DOSIMETRY OF I125 SEEDS [J].
HERBOLD, G ;
HARTMANN, G ;
TREUER, H ;
LORENZ, WJ .
PHYSICS IN MEDICINE AND BIOLOGY, 1988, 33 (09) :1037-1053
[8]  
IAEA, 1987, IAEA TECHN REP SER, V277
[9]   DOSIMETRIC CALCULATIONS AND MEASUREMENTS OF GOLD PLAQUE OPHTHALMIC IRRADIATORS USING IR-192 AND I-125 SEEDS [J].
LUXTON, G ;
ASTRAHAN, MA ;
LIGGETT, PE ;
NEBLETT, DL ;
COHEN, DM ;
PETROVICH, Z .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1988, 15 (01) :167-176
[10]  
MCLAUGHLIN WL, 1965, PHOTOGR SCI ENG, V9, P159