Results with the M. E. Muller cemented, straight-stem total hip prosthesis - A 10-year historical cohort study in 180 women

被引:20
作者
Havinga, ME [1 ]
Spruit, M [1 ]
Anderson, PG [1 ]
van Dijk-van Dam, MSW [1 ]
Pavlov, PW [1 ]
van Limbeek, J [1 ]
机构
[1] Sint Maartensklin, Dept Orthopaed, NL-6500 GM Nijmegen, Netherlands
关键词
MEM total hip prosthesis; survival rate; revision; follow-up;
D O I
10.1054/arth.2001.19003
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
In 1986. 242 M. E. Muller (MEM) cemented, straight-stem total hip arthroplasty prostheses were implanted in 229 patients: 15 hips (13 patients) were lost to follow-up. Of the remaining 227 implants. 180 were placed in women, and 47 were placed in men (each with mean age. 71 +/- 7.7 years). After 10 years. 66 patients had died, and 152 implants were still in situ. As a result of aseptic loosening, 9 hips were revised (5 femoral and 4 acetabular components): two of these patients had a Girdlestone as a result of postoperative infection. Because 50% of the men died during follow-up, further analysis was performed with the 180 implants in women. The incidence of revision for aseptic loosening was 5.9 per 1,000 implants. The cumulative survival rate after 10 years was 94%. Survival was nor influenced significantly by age, indication for operation, or having a contralateral hip prosthesis. The 10-year follow-up results for the MEM straight-stern total hip prosthesis in our hospital are satisfactory despite the probability that the cement mantle produced with this stem is not uniform in thickness.
引用
收藏
页码:33 / 36
页数:4
相关论文
共 18 条
[1]
BIEMANT JJ, 1995, REV CHIR ORTHOP REPA, V81, P380
[2]
BRAND RA, 1985, CLIN ORTHOP RELAT R, P237
[3]
THE EFFECT OF THE TYPE OF CEMENT ON EARLY REVISION OF CHARNLEY TOTAL HIP PROSTHESES - A REVIEW OF 8,579 PRIMARY ARTHROPLASTIES FROM THE NORWEGIAN ARTHROPLASTY REGISTER [J].
HAVELIN, LI ;
ESPEHAUG, B ;
VOLLSET, SE ;
ENGESAETER, LB .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1995, 77 (10) :1543-1550
[4]
Herberts P, 1997, CLIN ORTHOP RELAT R, P44
[5]
PLAIN RADIOGRAPHS INADEQUATE FOR EVALUATION OF THE CEMENT-BONE INTERFACE IN THE HIP-PROSTHESIS - A CADAVER STUDY OF FEMORAL STEMS [J].
JACOBS, ME ;
KOEWEIDEN, EMJ ;
SLOOFF, TJJH ;
HUISKES, R ;
VANHORN, JR .
ACTA ORTHOPAEDICA SCANDINAVICA, 1989, 60 (05) :541-543
[6]
Kohler A, 1997, Swiss Surg, V3, P44
[7]
ASEPTIC LOOSENING OF STRAIGHT-STEM AND CURVED-STEM MULLER FEMORAL PROSTHESES [J].
KRISMER, M ;
KLAR, M ;
KLESTIL, T ;
FRISCHHUT, B .
ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 1991, 110 (04) :190-194
[8]
PROGNOSIS OF TOTAL HIP-REPLACEMENT IN SWEDEN - FOLLOW-UP OF 92,675 OPERATIONS PERFORMED 1978-1990 [J].
MALCHAU, H ;
HERBERTS, P ;
AHNFELT, L .
ACTA ORTHOPAEDICA SCANDINAVICA, 1993, 64 (05) :497-506
[9]
MALCHAU H, 1998, 65 ANN M AM AC ORTH