Determination of a cyclic hexapeptide, a novel antifungal agent, in human plasma by high-performance liquid chromatography with ion spray and turbo ion spray tandem mass spectrometric detection

被引:28
作者
Chavez-Eng, CM [1 ]
Schwartz, M [1 ]
Constanzer, ML [1 ]
Matuszewski, BK [1 ]
机构
[1] Merck & Co Inc, Merck Res Labs, W Point, PA 19486 USA
来源
JOURNAL OF CHROMATOGRAPHY B | 1999年 / 721卷 / 02期
关键词
MK-0091;
D O I
10.1016/S0378-4347(98)00447-2
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Methods for the determination of a semi-synthetic cyclic hexapeptide (I, MK-0991) in human plasma based on high-performance liquid chromatography (HPLC) with tandem mass spectrometric (MS-MS) detection using pneumatically assisted electrospray (ion spray, ISP) and turbo ion spray (TISP) interfaces were developed. Drug and internal standard (II, an isostere of I) were isolated from plasma by solid-phase extraction (SPE). The eluent from SPE was evaporated to dryness, the residue was reconstituted in mobile phase and injected into the HPLC system. The use of ISP, TISP and heated nebulizer (HN) interfaces as sample introduction systems were evaluated and showed that the heated nebulizer was not adequate for analysis due to thermal instability and/or adsorption of I and II to glass surfaces of the interface. Compounds I and II were chromatographed on a wide pore (300 Angstrom), 150x4.6 mm C-8 analytical column, and the HPLC how-rate of 1.2 ml/min was split 1:20 prior to introduction to the ISP or TISP interface of the mass spectrometric system. The MS-MS detection was performed on a PE Sciex API III Plus tandem mass spectrometer operated in selected reaction monitoring mode (SRM). The precursor-->product ion combinations of mit 1093.7-->1033.6 and 1094.7-->1033.6 were used to quantify I and II, respectively, after chromatographic separation of the analytes. The assay was validated in the concentration range of 10-1000 ng/ml using ISP, and 2.5-500 ng/ml of plasma using TISP with good precision and adequate accuracy. The effects of HPLC mobile-phase components on the ionization efficiency and sensitivity of detection in the positive ionization mode, the evaluation of the matrix effect, and limitations in sensitivity of detection of I due to the formation of multiply charged species are presented. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:229 / 238
页数:10
相关论文
共 17 条
  • [1] Evaluation of the echinocandin antifungal MK-0991 (L-743,872): Efficacies in mouse models of disseminated aspergillosis, candidiasis, and cryptococcosis
    Abruzzo, GK
    Flattery, AM
    Gill, CJ
    Kong, L
    Smith, JG
    Pikounis, VB
    Balkovec, JM
    Bouffard, AF
    Dropinski, JF
    Rosen, H
    Kropp, H
    Bartizal, K
    [J]. ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1997, 41 (11) : 2333 - 2338
  • [2] ENHANCED SENSITIVITY FOR PEPTIDE-MAPPING WITH ELECTROSPRAY LIQUID-CHROMATOGRAPHY MASS-SPECTROMETRY IN THE PRESENCE OF SIGNAL SUPPRESSION DUE TO TRIFLUOROACETIC ACID-CONTAINING MOBILE PHASES
    APFFEL, A
    FISCHER, S
    GOLDBERG, G
    GOODLEY, PC
    KUHLMANN, FE
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1995, 712 (01) : 177 - 190
  • [3] In vitro preclinical evaluation studies with the echinocandin antifungal MK-0991 (L-743,872)
    Bartizal, K
    Gill, CJ
    Abruzzo, GK
    Flattery, AM
    Kong, L
    Scott, PM
    Smith, JG
    Leighton, CE
    Bouffard, A
    Dropinski, JF
    Balkovec, J
    [J]. ANTIMICROBIAL AGENTS AND CHEMOTHERAPY, 1997, 41 (11) : 2326 - 2332
  • [4] BARTIZAL K, 1993, CUTANEOUS ANTIFUNGAL, pCH29
  • [5] DETERMINATION OF NANOGRAM LEVELS OF PEPTIDE DRUG IN RABBIT AND HUMAN PLASMA USING HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COUPLED WITH ELECTROSPRAY-IONIZATION MASS-SPECTROMETRY
    CROWTHER, J
    ADUSUMALLI, V
    MUKHERJEE, T
    JORDAN, K
    ABUAF, P
    CORKUM, N
    GOLDSTEIN, G
    TOLAN, J
    [J]. ANALYTICAL CHEMISTRY, 1994, 66 (14) : 2356 - 2361
  • [6] MASS-SPECTROMETRIC QUANTIFICATION OF ENDOGENOUS BETA-ENDORPHIN
    DASS, C
    KUSMIERZ, JJ
    DESIDERIO, DM
    [J]. BIOLOGICAL MASS SPECTROMETRY, 1991, 20 (03) : 130 - 138
  • [7] MASS-SPECTROMETRIC ANALYSIS OF OPIOID AND TACHYKININ NEUROPEPTIDES IN NONSECRETING AND ACTH-SECRETING HUMAN PITUITARY-ADENOMAS
    DESIDERIO, DM
    KUSMIERZ, JJ
    ZHU, XG
    DASS, C
    HILTON, D
    ROBERTSON, JT
    SACKS, HS
    [J]. BIOLOGICAL MASS SPECTROMETRY, 1993, 22 (01) : 89 - 97
  • [8] FLATTERY A, 1996, INT C ANT AG CHEM 36
  • [9] QUANTITATIVE-ANALYSIS BY HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY ATMOSPHERIC-PRESSURE CHEMICAL IONIZATION MASS-SPECTROMETRY - THE DETERMINATION OF THE RENIN INHIBITOR CP-80,794 IN HUMAN SERUM
    FOUDA, H
    NOCERINI, M
    SCHNEIDER, R
    GEDUTIS, C
    [J]. JOURNAL OF THE AMERICAN SOCIETY FOR MASS SPECTROMETRY, 1991, 2 (02) : 164 - 167
  • [10] FU I, IN PRESS J PHARM BIO