Application of cloud-point extraction-reversed-phase high-performance liquid chromatography - A preliminary study of the extraction and quantification of vitamins A and E in human serum and whole blood

被引:80
作者
Sirimanne, SR
Patterson, DG
Ma, L
Justice, JB
机构
[1] Ctr Dis Control & Prevent, Div Environm Hlth Lab Sci, Natl Ctr Environm Hlth, Publ Hlth Serv,US Dept Hlth & Human Serv, Atlanta, GA 30341 USA
[2] Emory Univ, Dept Chem, Atlanta, GA 30322 USA
来源
JOURNAL OF CHROMATOGRAPHY B-ANALYTICAL TECHNOLOGIES IN THE BIOMEDICAL AND LIFE SCIENCES | 1998年 / 716卷 / 1-2期
关键词
cloud-point extraction; vitamins; retinol; alpha-tocopherol;
D O I
10.1016/S0378-4347(98)00287-4
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Methods available for quantification of vitamins A and E in serum or blood requires preconcentration and clean-up by liquid-liquid extraction, evaporation of the extract, and reconstitution of the extract in a solvent of choice before analysis. This process not only involves the use of toxic organic solvents but also requires a long sample preparation time. The lipids and other non-polar coextractants often require additional steps for sample clean-up and evaporation, which may cause sample losses. The use of cloud-point extraction eliminates most of these sample clean-up problems. We recently demonstrated that cloud-point extraction (CPE) can be used for extraction and quantification of polycyclic aromatic hydrocarbons (PAHs) and polychlorinated dibenzo-p-dioxins (PCDDs) from human serum. We now demonstrate how CPE can be used with human serum and blood, at volumes as low as 50 mu l, and report a methodology for extracting and quantifying two clinically important vitamins, (A and E) from human serum and blood. Vitamins A and E were extracted from human serum and blood by using Genapol X-80 as the cloud-point extractant under salting out conditions. Serum and blood samples were diluted in organic-free water to get sufficiently large sample volumes for CPE. The surfactant-rich phases were separated by centrifugation, and the samples were analyzed by HPLC-UV after deleterious coextractants were removed by precipitating them with acetonitrile. The recoveries of spiked vitamins A and E were found to be 85.6+/-0.4% and 82.6+/-5.2%, respectively. The average concentration of vitamins A and E in a serum pool after correction for recoveries were found to be 43.4+/-1.8 mu g/dl (1.5 +/- 0.1 mu mol/l) and 564.3 +/- 65.3 mu g/dl (13.1 +/- 1.5 mu mol/l), respectively. Vitamin A and E concentrations in whole blood were found to be 26.3+/-0.4 mu g/dl (0.92+/-0.01 mu mol/l) and 457.5+/-15.6 mu g/dl (10.6+/-0.4 mu mol/l), respectively. These values are comparable with those obtained by the reference method used at the Centers for Disease Control and Prevention. The success of the preliminary study will lead to a comprehensive validation of this method for vitamins A and E in serum and blood. (C) 1998 Elsevier Science BN. All rights reserved.
引用
收藏
页码:129 / 137
页数:9
相关论文
共 25 条
  • [1] ALFRED PA, 1994, J CHROMATOGR, V659, P289
  • [2] COMBINATION OF MICELLAR EXTRACTION OF POLYCYCLIC AROMATIC-HYDROCARBONS FROM AQUEOUS-MEDIA WITH DETECTION BY SYNCHRONOUS FLUORESCENCE
    BOCKELEN, A
    NIESSNER, R
    [J]. FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1993, 346 (04): : 435 - 440
  • [3] CLOUD POINT METHODOLOGY - A NEW APPROACH FOR PRECONCENTRATION AND SEPARATION IN HYDRODYNAMIC SYSTEMS OF ANALYSIS
    CORDERO, BM
    PAVON, JLP
    PINTO, CG
    LAESPADA, EF
    [J]. TALANTA, 1993, 40 (11) : 1703 - 1710
  • [4] GARCIA AL, 1992, ANAL CHIM ACTA, V264, P241
  • [5] STABILITY OF VITAMIN-E IN LONG-TERM STORED SERUM
    GUNTER, EW
    DRISKELL, WJ
    YEAGER, PR
    [J]. CLINICA CHIMICA ACTA, 1988, 175 (03) : 329 - 335
  • [6] A CRITICAL-REVIEW OF SURFACTANT-MEDIATED PHASE SEPARATIONS (CLOUD-POINT EXTRACTIONS) - THEORY AND APPLICATIONS
    HINZE, WL
    PRAMAURO, E
    [J]. CRITICAL REVIEWS IN ANALYTICAL CHEMISTRY, 1993, 24 (02) : 133 - 177
  • [7] SALTING-OUT SURFACTANT EXTRACTION OF PORPHYRINS AND METALLOPORPHYRIN FROM AQUEOUS NONIONIC SURFACTANT SOLUTIONS
    HORVATH, WJ
    HUIE, CW
    [J]. TALANTA, 1992, 39 (05) : 487 - 492
  • [8] KARK JD, 1981, J NATL CANCER I, V66, P7
  • [9] HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC METHOD FOR ROUTINE DETERMINATION OF VITAMIN-A AND VITAMIN-E AND BETA-CAROTENE IN PLASMA
    LEE, BL
    CHUA, SC
    ONG, HY
    ONG, CN
    [J]. JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS, 1992, 581 (01): : 41 - 47
  • [10] MA L, 1997, THESIS