Rapid slurry atomization using transverse heated electrothermal atomic absorption spectrometry for the determination of cadmium, copper, manganese and lead in biological reference materials

被引:30
作者
Meeravali, NN [1 ]
Kumar, SJ [1 ]
机构
[1] Natl Ctr Composit Characterizat Mat, Hyderabad 500062, Andhra Pradesh, India
关键词
transverse heated electrothermal atomic absorption spectrometry; rapid atomization; slurry sampling; cadmium; copper; manganese; lead; biological reference material;
D O I
10.1039/a801191e
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A rapid slurry atomization procedure for the determination of Cd, Cu, Mn and Pb in certified reference materials such as GBW 08504 Cabbage, NIST SRM 1566a Oyster Tissue and NIST SRM 1571 Orchard Leaves is described using transverse heated electrothermal atomic absorption spectrometry. The slurry was prepared in 5% HNO3 using a magnetic stirrer. No chemical modifier was used and the ashing step in the temperature programme was omitted. The atomization cycle was completed within 35-50 s. For Cd, Mn and Pb, aqueous calibration was used. For Cu, standard additions calibration was used. The Student's t-test at the 95% confidence level was carried out on the results obtained and good agreement was observed as compared with the certified values. Stability tests on slurries in glycerol and HNO3 indicated loss of Cd in glycerol in Cabbage as well as Oyster Tissue slurries. Pb was not stable in Cabbage slurry in 0.5% HNO3 medium and Cu was not stable in Oyster Tissue slurry in glycerol. The limits of detection obtained were 0.0056, 0.066, 0.068 and 0.02 mg kg(-1) for Cd, Pb, Cu and Mn, respectively, The precision of the results obtained varied between 0.2 and 2% using a rapid atomization programme.
引用
收藏
页码:647 / 652
页数:6
相关论文
共 20 条
[1]   THE INFLUENCE OF PYROLYSIS AND MATRIX MODIFIERS FOR ANALYSIS OF GLASS MATERIALS BY GFAAS USING SLURRY SAMPLE INTRODUCTION [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1990, 45 (07) :679-693
[2]   SOLID SAMPLING IN ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING COMMERCIAL ATOMIZERS - A REVIEW [J].
BENDICHO, C ;
DELOOSVOLLEBREGT, MTC .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1991, 6 (05) :353-374
[3]   Slurry sampling electrothermal atomic absorption spectrometric determination of lead, cadmium and manganese in human hair samples using rapid atomizer programs [J].
BermejoBarrera, P ;
MoredaPineiro, A ;
MoredaPineiro, J ;
BermejoBarrera, A .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (03) :301-306
[4]   RAPID SLURRY ANALYSIS OF SOLID COAL AND FLY-ASH SAMPLES [J].
BRADSHAW, D ;
SLAVIN, W .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1989, 44 (12) :1245-1256
[5]   SPEEDING UP DETERMINATIONS BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
HALLS, DJ .
ANALYST, 1984, 109 (08) :1081-1084
[6]   AUTOMATED TRACE-METAL ANALYSES OF SLURRIED SOLID SAMPLES BY GRAPHITE-FURNACE ATOMIC-ABSORPTION SPECTROMETRY WITH APPLICATION TO SEDIMENTS AND SUSPENDED MATTER COLLECTED IN NATURAL-WATERS [J].
HOENIG, M ;
REGNIER, P ;
WOLLAST, R .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1989, 4 (07) :631-634
[7]   Determination of silicon in biological tissue by electrothermal atomic absorption spectrometry using slurry sampling of original and pre-ashed samples [J].
Hornung, M ;
Krivan, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (10) :1123-1130
[8]  
JORDAN P, 1989, ATOM SPECTROSC, V10, P165
[9]   Slurry sampling device for use in electrothermal atomic absorption spectrometry [J].
LopezGarcia, I ;
SanchezMerlos, M ;
HernandezCordoba, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1997, 12 (07) :777-779
[10]   Rapid determination of selenium in soils and sediments using slurry sampling-electrothermal atomic absorption spectrometry [J].
LopezGarcia, I ;
SanchezMerlos, M ;
HernandezCordoba, M .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1996, 11 (10) :1003-1006