Fiducial point placement and the accuracy of point-based, rigid body registration

被引:312
作者
West, JB
Fitzpatrick, JM
Toms, SA
Maurer, CR
Maciunas, RJ
机构
[1] Vanderbilt Univ, Dept Elect Engn & Comp Sci, Nashville, TN 37235 USA
[2] Vanderbilt Univ, Med Ctr, Dept Neurol Surg, Nashville, TN 37235 USA
[3] Univ Rochester, Dept Neurol Surg, Rochester, NY USA
关键词
fiducial markers; point-based registration; target registration error;
D O I
10.1097/00006123-200104000-00023
中图分类号
R74 [神经病学与精神病学];
学科分类号
摘要
OBJECTIVE: To demonstrate that the shape of the configuration of fiducial points is an important factor governing target registration error (TRE) in point-based, rigid registration. METHODS: We consider two clinical situations: cranial neurosurgery and pedicle screw placement. For cranial neurosurgery, we apply theoretical results concerning TRE prediction, which we have previously derived and validated, to three hypothetical fiducial marker configurations. We illustrate the profile of expected TRE for each configuration. For pedicle screw placement, we apply the same theory to a common anatomic landmark configuration (tips of spinous and transverse processes) used for pedicle screw placement, and we estimate the error rate expected in placement of the screw. RESULTS: in the cranial neurosurgery example, we demonstrate that relatively small values of TRE may be achieved by using widely spread fiducial markers and/or placing the centroid of the markers near the target. We also demonstrate that near-collinear marker configurations far from the target may result in large TRE values. In the pedicle screw placement example, we demonstrate that the screw must be approximately 4 mm narrower than the pedicle in which it is implanted to minimize the chance of pedicle violation during placement. CONCLUSION: The placement of fiducial points is an important factor in minimizing the error rate for point-based, rigid registration. By using as many points as possible, avoiding near-collinear configurations, and ensuring that the centroid of the fiducial points is as near as possible to the target, TREs can be minimized.
引用
收藏
页码:810 / 816
页数:7
相关论文
共 18 条
[1]   Frameless stereotaxy with scalp-applied fiducial markers for brain biopsy procedures: experience in 218 cases [J].
Barnett, GH ;
Miller, DW ;
Weisenberger, J .
JOURNAL OF NEUROSURGERY, 1999, 91 (04) :569-576
[2]  
Darabi K, 1997, INT CONGR SER, V1134, P920
[3]  
ERICKSON BJ, 1993, AM J NEURORADIOL, V14, P713
[4]   Predicting error in rigid-body point-based registration [J].
Fitzpatrick, JM ;
West, JB ;
Maurer, CR .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1998, 17 (05) :694-702
[5]  
FITZPATRICK JM, 2000, MED IMAGE PROCESSING, V2, P447
[6]   REGISTRATION OF 3-D HEAD SURFACES USING MULTIPLE LANDMARKS [J].
FRIGHT, WR ;
LINNEY, AD .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1993, 12 (03) :515-520
[7]   COMBINED AND 3-DIMENSIONAL RENDERED MULTIMODAL DATA FOR PLANNING CRANIAL BASE SURGERY - A PROSPECTIVE EVALUATION [J].
GANDHE, AJ ;
HILL, DLG ;
STUDHOLME, C ;
HAWKES, DJ ;
RUFF, CF ;
COX, TCS ;
GLEESON, MJ ;
STRONG, AJ .
NEUROSURGERY, 1994, 35 (03) :463-470
[8]   Measurement of intraoperative brain surface deformation under a craniotomy [J].
Hill, DLG ;
Maurer, CR ;
Maciunas, RJ ;
Barwise, JA ;
Fitzpatrick, JM ;
Wang, MY .
NEUROSURGERY, 1998, 43 (03) :514-526
[9]   ACCURATE FRAMELESS REGISTRATION OF MR AND CT IMAGES OF THE HEAD - APPLICATIONS IN PLANNING SURGERY AND RADIATION-THERAPY [J].
HILL, DLG ;
HAWKES, DJ ;
GLEESON, MJ ;
COX, TCS ;
STRONG, AJ ;
WONG, WL ;
RUFF, CF ;
KITCHEN, ND ;
THOMAS, DGT ;
SOFAT, A ;
CROSSMAN, JE ;
STUDHOLME, C ;
GANDHE, AJ ;
GREEN, SEM ;
ROBINSON, GP .
RADIOLOGY, 1994, 191 (02) :447-454
[10]   Registration of head volume images using implantable fiducial markers [J].
Maurer, CR ;
Fitzpatrick, JM ;
Wang, MY ;
Galloway, RL ;
Maciunas, RJ ;
Allen, GS .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1997, 16 (04) :447-462