Tungsten-rhodium permanent chemical modifier for cadmium determination in fish slurries by electrothermal atomic absorption spectrometry

被引:71
作者
Lima, ÉC [1 ]
Krug, FJ [1 ]
Ferreira, AT [1 ]
Barbosa, F [1 ]
机构
[1] Univ Sao Paulo, Ctr Energia Nucl Agr, BR-13400970 Piracicaba, SP, Brazil
关键词
D O I
10.1039/a806645k
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A tungsten carbide-rhodium coating on the integrated platform of a transversely heated graphite atomizer was used as a permanent chemical modifier for the determination of Cd in fish slurries by electrothermal atomic absorption spectrometry. Slurries were sonicated for 25 s before being delivered to the previously W-Rh treated platform. The number of fish particles introduced into the atomizer for delivery of a 20 mu l slurry aliquot ranged from 7900 to 47 000. It was demonstrated that coating with 250 mu g of W + 200 mu g of Rh is as efficient as a Pd + Mg(NO3)(2) conventional modifier for obtaining good Cd recoveries (93.4-103.5%). The permanent W-Rh modifier remains stable for approximately 300 firings when 20 mu l of slurries containing up to 1.5% m/v are delivered into the atomizer. In addition, the permanent modifier increases the tube lifetime by 50-90% compared with untreated integrated platforms. Also, there is less degradation of sensitivity during the atomizer lifetime when compared with the conventional modifiers, resulting in a decreased need for re-calibration during routine analysis. The detection limit based on integrated absorbance was 1.4 ng g(-1) Cd for 1.50% m/v slurries. The RSD for 650 consecutive measurements of 20 mu l of 0.50% m/v fish slurry was 5.0% for the proposed permanent chemical modifier. Results for the determination of cadmium in fish slurries using the W + Rh permanent modifier were in agreement with those obtained with digested solutions by using Pd + Mg(NO3)(2), since no statistical differences were found by a paired t-test at the 99% level.
引用
收藏
页码:269 / 274
页数:6
相关论文
共 53 条
[1]  
BARRERA PB, 1994, J ANAL ATOM SPECTROM, V9, P469
[2]   Noble metals as permanent modifiers for the determination of mercury by electrothermal atomic absorption spectrometry [J].
Bulska, E ;
Kandler, W ;
Hulanicki, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (9-10) :1263-1270
[3]  
Cabrera C, 1995, J AOAC INT, V78, P1061
[4]  
*COMM SPECTR OTH O, 1978, SPECTROCHIM ACTA B, V33, P241
[5]   Evaluation of electrochemical hydride generation for the determination of arsenic and selenium in sea water by graphite furnace atomic absorption with in situ concentration [J].
Ding, WW ;
Sturgeon, RE .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1996, 51 (11) :1325-1334
[6]   DIRECT DETERMINATION OF IMPURITIES IN POWDERED SILICON-CARBIDE BY ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY USING THE SLURRY SAMPLING TECHNIQUE [J].
DOCEKAL, B ;
KRIVAN, V .
JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (03) :521-528
[7]  
DOIDGE PS, 1989, J ANAL ATOM SPECTROM, V4, P251, DOI 10.1039/ja9890400251
[8]   Determination of trace impurities in titanium dioxide by slurry sampling electrothermal atomic absorption spectrometry [J].
Dong, HM ;
Krivan, V ;
Welz, B ;
Schlemmer, G .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1997, 52 (12) :1747-1762
[9]   ANALYSIS OF SILICON-NITRIDE POWDERS FOR AL, CR, CU, PA, K, MG, MN, NA, AND ZN BY SLURRY-SAMPLING ELECTROTHERMAL ATOMIC-ABSORPTION SPECTROMETRY [J].
FRIESE, KC ;
KRIVAN, V .
ANALYTICAL CHEMISTRY, 1995, 67 (02) :354-359
[10]   Simultaneous determination of Se and As by hydride generation atomic absorption spectrometry with analyte concentration in a graphite furnace coated with zirconium [J].
Garbos, S ;
Walcerz, M ;
Bulska, E ;
Hulanicki, A .
SPECTROCHIMICA ACTA PART B-ATOMIC SPECTROSCOPY, 1995, 50 (13) :1669-1677