Radiation detection measurements with a new "Buried Junction" silicon avalanche photodiode

被引:44
作者
Lecomte, R
Pepin, C
Rouleau, D
McIntyre, RJ
McSween, D
Webb, P
机构
[1] Univ Sherbrooke, Dept Med Nucl & Radiobiol, Sherbrooke, PQ J1H 5N4, Canada
[2] EG&G Optoelect, Vaudreuil, PQ, Canada
关键词
avalanche photodiode; scintillation detector;
D O I
10.1016/S0168-9002(98)01182-6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
An improved version of a recently developed "Buried Junction" avalanche photodiode (APD), designed for use with scintillators. is described and characterized. This device. also called the "Reverse APD", is designed to have a wide depletion layer and thus low capacitance, but to have high gain only for e-h pairs generated within the first few microns of the depletion layer. Thus it has high gain for light from scintillators emitting in the 400 600 nm range, with relatively low dark current noise and it is relatively insensitive to minimum ionizing particles (MIPs). An additional feature is that the metallurgical junction is at the back of the wafer, leaving the front surface free to be coupled to a scintillator without fear of junction contamination. The modifications made in this device, as compared with the earlier diode, have resulted in a lower excess noise factor, lower dark current, and much-reduced trapping. The electrical and optical characteristics of this device are described and measurements of energy and timing resolution of this device with several scintillators (BGO, LSO and GSO) of potential interest in high-energy physics and PET imaging systems are presented. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:92 / 102
页数:11
相关论文
共 10 条
[1]  
ANDREACO MS, COMMUNICAITON
[3]   Non-proportionality in the scintillation response and the energy resolution obtainable with scintillation crystals [J].
Dorenbos, P ;
deHaas, JTM ;
vanEijk, CWE .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1995, 42 (06) :2190-2202
[4]   Investigation of GSO, LSO and YSO scintillators using reverse avalanche photodiodes [J].
Lecomte, R ;
Pepin, C ;
Rouleau, D ;
Saoudi, A ;
Andreaco, MS ;
Casey, M ;
Nutt, P ;
Dautet, H ;
Webb, PP .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1998, 45 (03) :478-482
[5]   STATUS OF BGO-AVALANCHE PHOTODIODE DETECTORS FOR SPECTROSCOPIC AND TIMING MEASUREMENTS [J].
LECOMTE, R ;
MARTEL, C ;
CARRIER, C .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1989, 278 (02) :585-597
[6]   A short-wavelength selective reach-through avalanche photodiode [J].
McIntyre, RJ ;
Webb, PP ;
Dautet, H .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1996, 43 (03) :1341-1346
[7]   ULTRA-LOW NOISE CHARGE SENSITIVE PREAMPLIFIER FOR SCINTILLATION DETECTION WITH AVALANCHE PHOTODIODES IN PET APPLICATIONS [J].
SCHMITT, D ;
LECOMTE, R ;
LAPOINTE, M ;
MARTEL, C ;
CARRIER, C ;
KARUTA, B ;
DUVAL, F .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1987, 34 (01) :91-96
[8]  
WEBB PP, 1974, RCA REV, V35, P234
[9]   LARGE AREA REACH-THROUGH AVALANCHE-DIODES FOR X-RAY SPECTROSCOPY [J].
WEBB, PP ;
MCINTYRE, RJ .
IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 1976, 23 (01) :138-144
[10]  
ZIEGLER S, COMMUNICAITON