Use of formalin-fixed, propidium iodide-stained human leukocytes as a standard for enumerating CD4+ T lymphocytes in a single-platform assay

被引:12
作者
Harrison, GM [1 ]
Bennett, AJ [1 ]
Moody, M [1 ]
Read, GF [1 ]
Williams, PE [1 ]
机构
[1] Univ Wales Hosp, Dept Med Biochem & Immunol, Cardiff CF14 4XW, S Glam, Wales
关键词
D O I
10.1128/CDLI.8.2.397-401.2001
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
A new, inexpensive method is described that enables lymphocytes to be enumerated very precisely. Normal leukocytes were simultaneously stained and fixed with a propidium iodide-paraformaldehyde solution. The preparation obtained (CellBeads) was used as an internal standard for cell enumeration by flow cytometry and was stable at 4 degreesC for at least 60 days. Unlike synthetic beads, the CellBeads behaved similarly to normal cells during red blood cell lysis and cell mashing procedures, When known numbers of CellBeads were added to whole blood samples and the numbers of CellBeads and lymphocytes were determined, highly reproducible and accurate enumerations were obtained-far more so than when synthetic beads were used. This inexpensive method is suitable for routine use.
引用
收藏
页码:397 / 401
页数:5
相关论文
共 12 条
[1]  
Barbosa IL, 1999, CYTOMETRY, V38, P274, DOI 10.1002/(SICI)1097-0320(19991215)38:6<274::AID-CYTO3>3.3.CO
[2]  
2-9
[3]  
Barnett D, 1999, BRIT J HAEMATOL, V106, P1059
[4]   Guidelines for the enumeration of CD4+ T lymphocytes in immunosuppressed individuals [J].
Barnett, D ;
Bird, G ;
Hodges, E ;
Linch, DC ;
Matutes, E ;
Newland, AC ;
Reilly, JT ;
Bain, BJ ;
Amos, R ;
Chapman, C ;
England, JM ;
Hyde, K .
CLINICAL AND LABORATORY HAEMATOLOGY, 1997, 19 (04) :231-241
[5]   MEASURING PERCENT LYMPHOCYTES BY FLOW-CYTOMETRY TO CALCULATE ABSOLUTE LYMPHOCYTE SUBSET COUNTS FOR HIV + SPECIMENS [J].
GALE, HB ;
HENRY, K .
CYTOMETRY, 1992, 13 (02) :175-181
[6]  
Lopez A, 1999, CYTOMETRY, V38, P231, DOI 10.1002/(SICI)1097-0320(19991015)38:5<231::AID-CYTO5>3.0.CO
[7]  
2-Y
[8]   IMMUNOPHENOTYPING SPECIMENS FROM HIV-INFECTED PERSONS - LABORATORY GUIDELINES FROM THE CENTERS-FOR-DISEASE-CONTROL AND PREVENTION [J].
NICHOLSON, JKA .
CYTOMETRY, 1994, 18 (01) :55-59
[9]   EFFECT OF TYPE OF HEMATOLOGY ANALYZER ON CD4 COUNT [J].
ROBINSON, G ;
MORGAN, L ;
EVANS, M ;
MCDERMOTT, S ;
PEREIRA, S ;
WANSBROUGHJONES, M ;
GRIFFIN, G .
LANCET, 1992, 340 (8817) :485-485
[10]   CD4+ T cell enumeration in HIV infection with limited resources [J].
Sherman, GG ;
Galpin, JS ;
Patel, JM ;
Mendelow, BV ;
Glencross, DK .