Comparative Evaluation of Three Automated Systems for DNA Extraction in Conjunction with Three Commercially Available Real-Time PCR Assays for Quantitation of Plasma Cytomegalovirus DNAemia in Allogeneic Stem Cell Transplant Recipients

被引:41
作者
Bravo, Dayana [1 ]
Angeles Clari, Maria [1 ]
Costa, Elisa [1 ]
Munoz-Cobo, Beatriz [1 ]
Solano, Carlos [2 ]
Jose Remigia, Maria [2 ]
Navarro, David [1 ,3 ]
机构
[1] Hosp Clin Univ, Microbiol Serv, Valencia 46010, Spain
[2] Hosp Clin Univ, Hematol & Med Oncol Serv, Valencia 46010, Spain
[3] Univ Valencia, Dept Microbiol, Sch Med, Valencia, Spain
关键词
AMPLICOR CMV MONITOR; PERIPHERAL-BLOOD; WHOLE-BLOOD; NESTED PCR; VIRAL LOAD; QUANTIFICATION; INFECTION; SPECIMENS; TESTS; KIT;
D O I
10.1128/JCM.00785-11
中图分类号
Q93 [微生物学];
学科分类号
071005 ; 100705 ;
摘要
Limited data are available on the performance of different automated extraction platforms and commercially available quantitative real-time PCR (QRT-PCR) methods for the quantitation of cytomegalovirus (CMV) DNA in plasma. We compared the performance characteristics of the Abbott mSample preparation system DNA kit on the m24 SP instrument (Abbott), the High Pure viral nucleic acid kit on the COBAS AmpliPrep system (Roche), and the EZ1 Virus 2.0 kit on the BioRobot EZ1 extraction platform (Qiagen) coupled with the Abbott CMV PCR kit, the LightCycler CMV Quant kit (Roche), and the Q-CMV complete kit (Nanogen), for both plasma specimens from allogeneic stem cell transplant (Allo-SCT) recipients (n = 42) and the OptiQuant CMV DNA panel (AcroMetrix). The EZ1 system displayed the highest extraction efficiency over a wide range of CMV plasma DNA loads, followed by the m24 and the AmpliPrep methods. The Nanogen PCR assay yielded higher mean CMV plasma DNA values than the Abbott and the Roche PCR assays, regardless of the platform used for DNA extraction. Overall, the effects of the extraction method and the QRT-PCR used on CMV plasma DNA load measurements were less pronounced for specimens with high CMV DNA content (> 10,000 copies/ml). The performance characteristics of the extraction methods and QRT-PCR assays evaluated herein for clinical samples were extensible at cell-based standards from AcroMetrix. In conclusion, different automated systems are not equally efficient for CMV DNA extraction from plasma specimens, and the plasma CMV DNA loads measured by commercially available QRT-PCRs can differ significantly. The above findings should be taken into consideration for the establishment of cutoff values for the initiation or cessation of preemptive antiviral therapies and for the interpretation of data from clinical studies in the Allo-SCT setting.
引用
收藏
页码:2899 / 2904
页数:6
相关论文
共 26 条
  • [1] Evaluation of an automated extraction system in combination with Affigene® CMV Trender for CMV DNA quantitative determination:: Comparison with nested PCR and pp65 antigen test
    Abbate, I.
    Finnstrom, N.
    Zaniratti, S.
    Solmone, M. C.
    Selvaggini, S.
    Bennici, E.
    Neri, S.
    Brega, C.
    Paterno, M.
    Capobianchi, M. R.
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2008, 151 (01) : 61 - 65
  • [2] Evaluation of a novel real-time PCR system for cytomegalovirus DNA quantitation on whole blood and correlation with pp65-antigen test in guiding pre-emptive antiviral treatment
    Allice, Tiziano
    Cerutti, Francesco
    Pittaluga, Fabrizia
    Varetto, Silvia
    Franchello, Alessandro
    Salizzoni, Mauro
    Ghisetti, Valeria
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2008, 148 (1-2) : 9 - 16
  • [3] [Anonymous], WHOBS102138 ECBS
  • [4] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [5] Human cytomegalovirus DNA in plasma and serum specimens of renal transplant recipients is highly fragmented
    Boom, R
    Sol, CJA
    Schuurman, T
    van Breda, A
    Weel, JFL
    Beld, M
    ten Berge, IJM
    Wertheim-van Dillen, PME
    de Jong, MD
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2002, 40 (11) : 4105 - 4113
  • [6] Comparison of quantitative and qualitative PCR assays for cytomegalovirus DNA in plasma
    Caliendo, AM
    Schuurman, R
    Yen-Lieberman, B
    Spector, SA
    Andersen, J
    Manjiry, R
    Crumpacker, C
    Lurain, NS
    Erice, A
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (04) : 1334 - 1338
  • [7] Evaluation of real-time PCR laboratory-developed tests using analyte-specific reagents for cytomegalovirus quantification
    Caliendo, Angela M.
    Ingersoll, Jessica
    Fox-Canale, Andrea M.
    Pargman, Sabine
    Bythwood, Tameka
    Hayden, Mary K.
    Bremer, James W.
    Lurain, Nell S.
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2007, 45 (06) : 1723 - 1727
  • [8] Real-time PCR in clinical microbiology: Applications for a routine laboratory testing
    Espy, MJ
    Uhl, JR
    Sloan, LM
    Buckwalter, SP
    Jones, MF
    Vetter, EA
    Yao, JDC
    Wengenack, NL
    Rosenblatt, JE
    Cockerill, FR
    Smith, TF
    [J]. CLINICAL MICROBIOLOGY REVIEWS, 2006, 19 (01) : 165 - +
  • [9] Quantification of DNA in plasma by an automated real-time PCR assay (cytomegalovirus PCR kit) for surveillance of active cytomegalovirus infection and guidance of preemptive therapy for allogeneic hematopoietic stem cell transplant recipients
    Gimeno, Concepcion
    Solano, Carlos
    Latorre, Jose C.
    Hernandez-Boluda, Juan C.
    Clari, Maria A.
    Remigia, Maria J.
    Furio, Santiago
    Calabuig, Marisa
    Tormo, Nuria
    Navarro, David
    [J]. JOURNAL OF CLINICAL MICROBIOLOGY, 2008, 46 (10) : 3311 - 3318
  • [10] Multicentric evaluation of a new commercial cytomegalovirus real-time PCR quantitation assay
    Gouarin, S.
    Vabret, A.
    Scieux, C.
    Agbalika, F.
    Cherot, J.
    Mengelle, C.
    Deback, C.
    Petitjean, J.
    Dina, J.
    Freymuth, F.
    [J]. JOURNAL OF VIROLOGICAL METHODS, 2007, 146 (1-2) : 147 - 154