A computational tool to support pre-operative planning of stentless aortic valve implant

被引:30
作者
Auricchio, F. [1 ,2 ,3 ]
Conti, M. [1 ]
Morganti, S. [1 ]
Totaro, P. [4 ]
机构
[1] Univ Pavia, DMS, I-27100 Pavia, Italy
[2] CNR, IMATI, I-27100 Pavia, Italy
[3] IUSS, CeSNA, Pavia, Italy
[4] IRCCS Policlin San Matteo, Dipartimento Chirurg Cardiotoracovasc, Pavia, Italy
关键词
Finite Element Analysis (FEA); Patient-specific model; Aortic valve replacement (AVR); REPLACEMENT; ROOT;
D O I
10.1016/j.medengphy.2011.05.006
中图分类号
R318 [生物医学工程];
学科分类号
100103 [病原生物学];
摘要
In some cases of aortic valve leaflet disease, the implant of a stentless biological prosthesis represents an excellent option for aortic valve replacement (AVR). In particular, if compared to more classical surgical approaches, it provides a more physiological hemodynamic performance and a minor trombogeneticity avoiding the use of anticoagulants. The clinical outcomes of AVR are strongly dependent on an appropriate choice of both prosthesis size and replacement technique, which are, at present, strictly related to surgeon's experience and skill. Therefore, also this treatment, like most reconstructive procedures in cardiac surgery, remains "more art than science" [1]. Nowadays computational methodologies represent a useful tool both to investigate the aortic valve behavior, in physiologic and pathologic conditions and to reproduce virtual post-operative scenarios. The present study aims at supporting the AVR procedure planning through a patient-specific Finite Element Analysis (FEA) of stentless valve implantation. Firstly, we perform FEA to simulate the prosthesis placement inside the patient-specific aortic root; then, we reproduce, again by means of FEA, the diastolic closure of the valve to evaluate both the coaptation and the stress/strain state. The simulation results prove that both the valve size and the anatomical asymmetry of the Valsalva sinuses affect the prosthesis placement procedure. (C) 2011 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1183 / 1192
页数:10
相关论文
共 32 条
[1]
Aboud A, 2010, ASIAN CARDIOVASCULAR, V17, P35
[2]
Finite element analysis of aortic root dilation: a new procedure to reproduce pathology based on experimental data [J].
Auricchio, F. ;
Conti, M. ;
Demertzis, S. ;
Morganti, S. .
COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2011, 14 (10) :875-882
[3]
The Sorin Freedom SOLO stentless aortic valve: Technique of implantation and operative results in 109 patients [J].
Aymard, Thierry ;
Eckstein, Friedrich ;
Englberger, Lars ;
Stalder, Mario ;
Kadner, Alexander ;
Carrel, Thierry .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2010, 139 (03) :775-777
[4]
Beholz Sven, 2006, Asian Cardiovasc Thorac Ann, V14, P128
[5]
Dynamic finite element analysis of the aortic root from MRI-derived parameters [J].
Conti, Carlo A. ;
Votta, Emiliano ;
Della Corte, Alessandro ;
Del Viscovo, Luca ;
Bancone, Ciro ;
Cotrufo, Maurizio ;
Redaelli, Alberto .
MEDICAL ENGINEERING & PHYSICS, 2010, 32 (02) :212-221
[6]
Aortic valve sparing operations [J].
David, TE .
ANNALS OF THORACIC SURGERY, 2002, 73 (04) :1029-1030
[7]
Aortic valve sparing operations: basic concepts [J].
Feindel, CM ;
David, TE .
INTERNATIONAL JOURNAL OF CARDIOLOGY, 2004, 97 :61-66
[8]
Ganguly Gautam, 2004, Heart Lung Circ, V13, P161, DOI 10.1016/j.hlc.2004.02.004
[9]
GLAUBER M, 2007, MULTIMEDIA MANUAL CA, V13, P161
[10]
Mechanisms of aortic valve incompetence: Finite-element modeling of Marfan syndrome [J].
Grande-Allen, KJ ;
Cochran, RP ;
Reinhall, PG ;
Kunzelman, KS .
JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2001, 122 (05) :946-954