A cobalt-carbon eutectic fixed point for the calibration of contact thermometers at temperatures above 1100°C

被引:51
作者
Edler, F [1 ]
Baratto, AC [1 ]
机构
[1] Phys Tech Bundesanstalt, D-1000 Berlin, Germany
关键词
D O I
10.1088/0026-1394/42/4/003
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A vertical cobalt-carbon (Co-C) eutectic fixed point cell was constructed at PTB to demonstrate its use for improvement of the calibration of noble-metal thermocouples at temperatures above 1100 degrees C. The melting and freezing temperatures of the Co-C eutectic were measured in different high-temperature furnaces at PTB and INMETRO (Brazil) to show its stability by using a Pt/Pd thermocouple. The reproducibility of all measured electromotive forces at the inflection points of the melting curves amounts to a value of about 0.06 degrees C. No drift in the melting temperature was observed. Therefore, the Co-C eutectic fixed point cell can be used as an adequate transfer standard for dissemination of the International Temperature Scale of 1990 (ITS-90) in the temperature range above the freezing point of copper.
引用
收藏
页码:201 / 207
页数:7
相关论文
共 11 条
[1]  
EDLER F, 1993, 935 CCT
[2]  
EDLER F, 2005, UNPUB P TEMPM 2004
[3]  
Jahan F, 2003, AIP CONF PROC, V684, P469, DOI 10.1063/1.1627170
[4]  
LOWE D, 2005, UNPUB P TEMPM 2004
[5]  
MORICE R, 2005, UNPUB P TEMPM 2004
[6]  
NICHOLAS JV, 2001, TRACEABLE TEMPERATUR, P135
[7]   THE INTERNATIONAL TEMPERATURE SCALE OF 1990 (ITS-90) [J].
PRESTONTHOMAS, H .
METROLOGIA, 1990, 27 (01) :3-10
[8]  
SADLI M, 2005, UNPUB P TEMPM 04
[9]   Radiometric observation of melting and freezing plateaus for a series of metal-carbon eutectic points in the, range 1330°C to 1950°C [J].
Yamada, Y ;
Sakate, H ;
Sakuma, F ;
Ono, A .
METROLOGIA, 1999, 36 (03) :207-209
[10]   Thermocouple observations of melting and freezing plateaus for metal-carbon eutectics between the copper and palladium points [J].
Yamada, Y ;
Sakuma, F ;
Ono, A .
METROLOGIA, 2000, 37 (01) :71-73