Characterization of wide-pore reversed-phase columns for biopolymer separations .1. Single-parametric evaluation

被引:9
作者
Bede, A
Rippel, G
Szepesy, L
Claessens, HA
机构
[1] TECH UNIV BUDAPEST,DEPT AGR CHEM TECHNOL,H-1521 BUDAPEST,HUNGARY
[2] TECH UNIV BUDAPEST,DEPT CHEM TECHNOL,H-1521 BUDAPEST,HUNGARY
[3] EINDHOVEN UNIV TECHNOL,INSTRUMENTAL ANAL LAB,5600 MB EINDHOVEN,NETHERLANDS
关键词
stationary phases; LC; hydrophobicity; proteins; amino acids; peptides; benzenes;
D O I
10.1016/0021-9673(95)01217-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Nine prepacked wide-pore columns recommended for protein separation were studied. The stationary phases investigated were widely different with respect to the type of ligands. In order to characterize the columns, measurements were made with six different test mixtures containing two different homologous series, substituted benzenes and amino acids, peptides and proteins. The isopotential eluent compositions obtained for the homologous series revealed the polar-apolar nature of the stationary phases, The relative hydrophobicity calculated for the individual test compounds provided more thorough but still partial characterization. The involvement of substituted benzenes resulted in the most complete description. The characterization with amino acids, peptides and proteins furnished results similar to those obtained for smaller test compounds having low or intermediate polarity. The examination of the behaviour of proteins revealed that the retention is determined by the hydrophobicity of the stationary phases and the other characteristics affect the selectivity.
引用
收藏
页码:179 / 188
页数:10
相关论文
共 36 条
  • [1] RETENTION MECHANISM FOR REVERSED-PHASE ION-PAIR LIQUID-CHROMATOGRAPHY
    BIDLINGMEYER, BA
    DEMING, SN
    PRICE, WP
    SACHOK, B
    PETRUSEK, M
    [J]. JOURNAL OF CHROMATOGRAPHY, 1979, 186 (DEC): : 419 - 434
  • [2] CHROMATOGRAPHIC AND SOLID-STATE NUCLEAR MAGNETIC-RESONANCE STUDY OF THE CHANGES IN REVERSED-PHASE PACKINGS FOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY AT DIFFERENT ELUENT COMPOSITIONS
    CLAESSENS, HA
    DEHAAN, JW
    VANDEVEN, LJM
    DEBRUYN, PC
    CRAMERS, CA
    [J]. JOURNAL OF CHROMATOGRAPHY, 1988, 436 (03): : 345 - 365
  • [3] CHEMOMETRIC CLASSIFICATION OF REVERSED-PHASE HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY COLUMNS
    DELANEY, MF
    PAPAS, AN
    WALTERS, MJ
    [J]. JOURNAL OF CHROMATOGRAPHY, 1987, 410 (01): : 31 - 41
  • [4] DIBUSSOLO JM, 1984, AM BIOTECHNOL LAB, V2, P1
  • [5] DONG M W, 1989, Biochromatography, V4, P19
  • [6] CHROMATOGRAPHIC CHARACTERIZATION OF SILICA-BASED REVERSED PHASES
    ENGELHARDT, H
    LOW, H
    GOTZINGER, W
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 544 (1-2): : 371 - 379
  • [7] COMPARISON AND CHARACTERIZATION OF REVERSED PHASES
    ENGELHARDT, H
    JUNGHEIM, M
    [J]. CHROMATOGRAPHIA, 1990, 29 (1-2) : 59 - 68
  • [8] CHROMATOGRAPHIC CHARACTERIZATION OF SILICA SURFACES
    ENGELHARDT, H
    MULLER, H
    [J]. JOURNAL OF CHROMATOGRAPHY, 1981, 218 (1-3): : 395 - 407
  • [9] PROPERTIES AND DIVERSITY OF C18 BONDED PHASES
    ENGELHARDT, H
    DREYER, B
    SCHMIDT, H
    [J]. CHROMATOGRAPHIA, 1982, 16 : 11 - 17
  • [10] COMPARISON OF COLUMNS FOR REVERSED-PHASE LIQUID-CHROMATOGRAPHY
    GOLDBERG, AP