摘要
OBJECTIVE, The purpose of our study was to establish the most suitable algorithm to compare coronary artery calcium measurements performed with electron beam CT and multi row detector CT for the assessment of coronary artery disease. SUBJECTS AND METHODS. Coronary artery screening was performed in 100 patients with both electron beam and multirow detector CT. The images were transferred to a dedicated workstation for determination of the calcium score, volume, mass, density, and number of lesions. III addition to the traditional threshold of 130 H, the score of multirow detector CT studies was reevaluated at a threshold of 90 H. Fifty-nine of the patients underwent conventional coronary catheterization. Receiver operating characteristic curve analysis of the different scoring algorithms for detection of significant coronary artery stenosis was performed. RESULTS. The correlation between electron beam CT and multirow detector CT was high for every quantification algorithm. Determination of the score and the number of lesions with multirow detector CT revealed a systematic error of the measurement compared with electron beam CT. The areas under the curve in the receiver operating characteristic curve analyses for electron beam and multirow detector CT were similar for the score. volume, and mass, whereas they were lower for the density. No significant difference was found for the areas under the curve between scores using a 130-H and those using a 90-H threshold. CONCLUSION. Volume and mass indexes are superior to the traditional score, density, and number of lesions for comparing the results of electron beam and multirow detector CT and for determining significant coronary artery disease.
引用
收藏
页码:1295 / 1298
页数:4
相关论文
共 24 条
  • [1] QUANTIFICATION OF CORONARY-ARTERY CALCIUM USING ULTRAFAST COMPUTED-TOMOGRAPHY
    AGATSTON, AS
    JANOWITZ, WR
    HILDNER, FJ
    ZUSMER, NR
    VIAMONTE, M
    DETRANO, R
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1990, 15 (04) : 827 - 832
  • [2] Helical and single-slice conventional CT versus electron beam CT for the quantification of coronary artery calcification
    Becker, CR
    Jakobs, TF
    Aydemir, S
    Becker, A
    Knez, A
    Schoepf, UJ
    Bruening, R
    Haberl, R
    Reiser, MF
    [J]. AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 174 (02) : 543 - 547
  • [3] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [4] Effect of HMG-Coa reductase inhibitors on coronary artery disease as assessed by electron-beam computed tomography
    Callister, TQ
    Raggi, P
    Cooil, B
    Lippolis, NJ
    Russo, DJ
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1998, 339 (27) : 1972 - 1978
  • [5] Coronary artery disease: Improved reproducibility of calcium scoring with an electron-beam CT volumetric method
    Callister, TQ
    Cooil, B
    Raya, SP
    Lippolis, NJ
    Russo, DJ
    Raggi, P
    [J]. RADIOLOGY, 1998, 208 (03) : 807 - 814
  • [6] Detrano R, 1995, Am J Card Imaging, V9, P167
  • [7] ACCURACY OF QUANTIFYING CORONARY HYDROXYAPATITE WITH ELECTRON-BEAM TOMOGRAPHY
    DETRANO, R
    KANG, XP
    MAHAISAVARIYA, P
    TANG, WY
    COLOMBO, A
    MOLLOI, S
    GARNER, D
    NICKERSON, S
    [J]. INVESTIGATIVE RADIOLOGY, 1994, 29 (08) : 733 - 738
  • [8] Electron-beam CT: The effect of using a correction function on coronary artery calcium quantitation
    Greaser, LE
    Yoon, HC
    Mather, RT
    McNitt-Gray, M
    Goldin, JG
    [J]. ACADEMIC RADIOLOGY, 1999, 6 (01) : 40 - 48
  • [9] A METHOD OF COMPARING THE AREAS UNDER RECEIVER OPERATING CHARACTERISTIC CURVES DERIVED FROM THE SAME CASES
    HANLEY, JA
    MCNEIL, BJ
    [J]. RADIOLOGY, 1983, 148 (03) : 839 - 843
  • [10] THE MEANING AND USE OF THE AREA UNDER A RECEIVER OPERATING CHARACTERISTIC (ROC) CURVE
    HANLEY, JA
    MCNEIL, BJ
    [J]. RADIOLOGY, 1982, 143 (01) : 29 - 36