摘要
The paper describes the current performance of the primary clock CS1 of the Physikalisch-Technische Bundesanstalt (PTB). The clock employs a thermal atomic beam and magnetic state selection. After major reconstruction during 1995 and 1996, routine clock operation was restarted in May 1997. Results of comparisons with UTC(PTB) and other internationally recognized time scales are presented. An evaluation of the CS1 Type B uncertainty yielded 7x10(-15) (1 sigma). The CS1 is thus the most accurate atomic clock in continuous operation world-wide.
引用
收藏
页码:829 / 845
页数:17
相关论文
共 34 条
[1]  
AUDOIN C, 1978, IEEE T INSTRUM MEAS, V27, P325, DOI 10.1109/TIM.1978.4314705
[2]  
AUGUSTIN R, 1994, P 11 EUR FREQ TIM FO, P47
[3]   Experimental verification of the shift of the cesium hyperfine transition frequency due to blackbody radiation [J].
Bauch, A ;
Schroder, R .
PHYSICAL REVIEW LETTERS, 1997, 78 (04) :622-625
[4]   The PTB primary clock CS3: Type B evaluation of its standard uncertainty [J].
Bauch, A ;
Heindorff, T ;
Schroder, R ;
Fischer, B .
METROLOGIA, 1996, 33 (03) :249-259
[5]   MEASUREMENT OF THE FREQUENCY-SHIFT DUE TO DISTRIBUTED CAVITY PHASE DIFFERENCE IN AN ATOMIC CLOCK [J].
BAUCH, A ;
HEINDORFF, T ;
SCHROEDER, R .
IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 1985, 34 (02) :136-138
[6]  
BAUCH A, 1993, ANN PHYS-LEIPZIG, V2, P421, DOI 10.1002/andp.19935050502
[7]  
BAUCH A, IN PRESS IEEE T INST
[8]  
BAUCH A, 1994, P 8 EUR FREQ TIM FOR, P503