A new variable dispersion double-focusing plasma mass spectrometer with performance illustrated for Pb isotopes

被引:298
作者
Belshaw, NS [1 ]
Freedman, PA
O'Nions, RK
Frank, M
Guo, Y
机构
[1] Univ Oxford, Dept Earth Sci, Oxford OX1 3PR, England
[2] Nu Instruments Ltd, Wrexham, Wales
关键词
plasma; mass spectrometer; multiple collector; isotopes; lead;
D O I
10.1016/S1387-3806(98)14150-7
中图分类号
O64 [物理化学(理论化学)、化学物理学]; O56 [分子物理学、原子物理学];
学科分类号
070203 ; 070304 ; 081704 ; 1406 ;
摘要
A new generation multiple collector plasma source mass spectrometer (PSMS) produced by Nu Instruments Ltd is evaluated. The instrument has a double-focusing Nier-Johnson analyser with laminated magnet and a novel variable dispersion ion optical arrangement, enabling all masses to be located in the centre of the Faraday collectors of a fixed static array. The performance of the instrument has been assessed through the analysis of NBS-981 Pb using a Tl doping technique with Faraday collector efficiencies and amplifier gains determined independently. A second method of analysis involves comparison of interspersed standard and sample ph measurements with effective gains for each collector determined from the standards. In both cases the repetition of ph isotope measurement is competetive with the thermal ionization mass spectrometry (TIMS) double-spike method. (Int J Mass Spectrom 181 (1998) 51-58) (C) 1998 Elsevier Science B.V.
引用
收藏
页码:51 / 58
页数:8
相关论文
共 12 条
  • [1] IN-SITU SR ISOTOPIC ANALYSIS BY LASER-ABLATION
    CHRISTENSEN, JN
    HALLIDAY, AN
    LEE, DC
    HALL, CM
    [J]. EARTH AND PLANETARY SCIENCE LETTERS, 1995, 136 (1-2) : 79 - 85
  • [2] Climate and ocean dynamics and the lead isotopic records in Pacific ferromanganese crusts
    Christensen, JN
    Halliday, AN
    Godfrey, LV
    Hein, JR
    Rea, DK
    [J]. SCIENCE, 1997, 277 (5328) : 913 - 918
  • [3] Applications of multiple collector-ICPMS to cosmochemistry, geochemistry, and paleoceanography
    Halliday, AN
    Lee, DC
    Christensen, JN
    Rehkamper, M
    Yi, W
    Luo, XZ
    Hall, CM
    Ballentine, CJ
    Pettke, T
    Stirling, C
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 1998, 62 (06) : 919 - 940
  • [4] Lead isotopic analyses of NIST standard reference materials using multiple collector inductively coupled plasma mass spectrometry coupled with a modified external correction method for mass discrimination effect
    Hirata, T
    [J]. ANALYST, 1996, 121 (10) : 1407 - 1411
  • [5] Evolution of Nd and Pb isotopes in Central Pacific seawater from ferromanganese crusts
    Ling, HF
    Burton, KW
    ONions, RK
    Kamber, BS
    von Blanckenburg, F
    Gibb, AJ
    Hein, JR
    [J]. EARTH AND PLANETARY SCIENCE LETTERS, 1997, 146 (1-2) : 1 - 12
  • [6] Precise and accurate neodymium isotopic measurements by plasma-source mass spectrometry
    Luais, B
    Telouk, P
    Albarede, F
    [J]. GEOCHIMICA ET COSMOCHIMICA ACTA, 1997, 61 (22) : 4847 - 4854
  • [7] VALIDATION OF THE ANALYTICAL LINEARITY AND MASS DISCRIMINATION CORRECTION MODEL EXHIBITED BY A MULTIPLE COLLECTOR INDUCTIVELY-COUPLED PLASMA-MASS SPECTROMETER BY MEANS OF A SET OF SYNTHETIC URANIUM ISOTOPE MIXTURES
    TAYLOR, PDP
    DEBIEVRE, P
    WALDER, AJ
    ENTWISTLE, A
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1995, 10 (05) : 395 - 398
  • [8] Todt W., 1996, GEOPHYS MONOGR SER, DOI [DOI 10.1029/GM095, 10.1029/GM095p0429, DOI 10.1029/GM095P0429]
  • [9] ISOTOPE RATIO MEASUREMENT OF LEAD, NEODYMIUM AND NEODYMIUM SAMARIUM MIXTURES, HAFNIUM AND HAFNIUM LUTETIUM MIXTURES WITH A DOUBLE FOCUSING MULTIPLE COLLECTOR INDUCTIVELY COUPLED PLASMA MASS-SPECTROMETER
    WALDER, AJ
    PLATZNER, I
    FREEDMAN, PA
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1993, 8 (01) : 19 - 23
  • [10] ISOTOPIC RATIO MEASUREMENT USING A DOUBLE FOCUSING MAGNETIC-SECTOR MASS ANALYZER WITH AN INDUCTIVELY COUPLED PLASMA AS AN ION-SOURCE
    WALDER, AJ
    FREEDMAN, PA
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 1992, 7 (03) : 571 - 575