Cardiovascular stent design and vessel stresses: a finite element analysis

被引:317
作者
Lally, C
Dolan, F
Prendergast, PJ [1 ]
机构
[1] Trinity Coll Dublin, Dept Mech Engn, Ctr Bioengn, Dublin 2, Ireland
[2] Medtron Vasc, Galway, Ireland
关键词
coronary stent; finite-element method; restenosis; vascular injury; arterial-wall mechanics;
D O I
10.1016/j.jbiomech.2004.07.022
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Intravascular stents of various designs are currently in use to restore patency in atherosclerotic coronary arteries and it has been found that different stents have different in-stent restenosis rates. It has been hypothesized that the level of vascular injury caused to a vessel by a stent determines the level of restenosis. Computational studies may be used to investigate the mechanical behaviour of stents and to determine the biomechanical interaction between the stent and the artery in a stenting procedure. In this paper, we test the hypothesis that two different stent designs will provoke different levels of stress within an atherosclerotic artery and hence cause different levels of vascular injury. The stents analysed using the finite-element method were the S7 (Medtronic AVE) and the NIR (Boston Scientific) stent designs. An analysis of the arterial wall stresses in the stented arteries indicates that the modular S7 stent design causes lower stress to an atherosclerotic vessel with a localized stenotic lesion compared to the slotted tube NIR design. These results correlate with observed clinical restenosis rates, which have found higher restenosis rates in the NIR compared with the S7 stent design. Therefore, the testing methodology outlined here is proposed as a pre-clinical testing tool, which could be used to compare and contrast existing stent designs and to develop novel stent designs. (c) 2004 Elsevier Ltd. All rights reserved.
引用
收藏
页码:1574 / 1581
页数:8
相关论文
共 35 条
[1]  
Arakawa K, 1998, J Cardiol, V32, P379
[2]  
Auricchio F., 2001, COMPUT METHOD BIOMEC, V4, P249, DOI DOI 10.1080/10255840108908007
[3]   Final results of a randomized trial comparing the NIR stent to the Palmaz-Schatz stent for narrowings in native coronary arteries [J].
Baim, DS ;
Cutlip, DE ;
O'Shaughnessy, CD ;
Hermiller, JB ;
Kereiakes, DJ ;
Giambartolomei, A ;
Katz, S ;
Lansky, AJ ;
Fitzpatrick, M ;
Popma, JJ ;
Ho, KKL ;
Leon, MB ;
Kuntz, RE .
AMERICAN JOURNAL OF CARDIOLOGY, 2001, 87 (02) :152-156
[4]   COMPRESSIBILITY OF ARTERIAL WALL [J].
CAREW, TE ;
VAISHNAV, RN ;
PATEL, DJ .
CIRCULATION RESEARCH, 1968, 23 (01) :61-&
[5]   Finite element simulation of stent and balloon interaction [J].
Chua, SND ;
Mac Donald, BJ ;
Hashmi, MSJ .
JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2003, 143 :591-597
[6]   Selection of coronary stents [J].
Colombo, A ;
Stankovic, G ;
Moses, JW .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2002, 40 (06) :1021-1033
[7]   INFLUENCE OF VASCULAR SMOOTH MUSCLE ON CONTRACTILE MECHANICS AND ELASTICITY OF ARTERIES [J].
DOBRIN, PB ;
ROVICK, AA .
AMERICAN JOURNAL OF PHYSIOLOGY, 1969, 217 (06) :1644-&
[8]   Mechanical behaviour modelling of balloon-expandable stents [J].
Dumoulin, C ;
Cochelin, B .
JOURNAL OF BIOMECHANICS, 2000, 33 (11) :1461-1470
[9]   Pathobiologic responses to stenting [J].
Edelman, ER ;
Rogers, C .
AMERICAN JOURNAL OF CARDIOLOGY, 1998, 81 (7A) :4E-6E
[10]   Mechanical properties of coronary stents determined by using finite element analysis [J].
Etave, F ;
Finet, G ;
Boivin, M ;
Boyer, JC ;
Rioufol, G ;
Thollet, G .
JOURNAL OF BIOMECHANICS, 2001, 34 (08) :1065-1075