Cutting thread at flexible endoscopy

被引:17
作者
Gong, F
Swain, P
Kadirkamanathan, S
Hepworth, C
Laufer, J
Shelton, J
Mills, T
机构
[1] ROYAL LONDON HOSP,GASTROINTESTINAL SCI RES UNIT,LONDON E1 1BB,ENGLAND
[2] UNIV LONDON QUEEN MARY & WESTFIELD COLL,IRC BIOMED MAT,DEPT MED PHYS & BIOENGN,LONDON E1 4NS,ENGLAND
基金
英国惠康基金;
关键词
D O I
10.1016/S0016-5107(96)70049-4
中图分类号
R57 [消化系及腹部疾病];
学科分类号
摘要
Background: New thread-cutting techniques were developed for use at flexible endoscopy. A guillotine was designed to follow and cut thread at the endoscope tip. A new method was developed for guiding suture cutters. Efficacy of Nd:YAG laser cutting of threads was studied. Experimental and clinical experience with thread-cutting methods is presented. Methods: A 2.4 mm diameter flexible thread-cutting guillotine was constructed featuring two lateral holes with sharp edges through which sutures to be cut are passed. Standard suture cutters were guided by backloading thread through the cutters extracorporeally. A snare cutter was constructed to retrieve objects sewn to tissue. Efficacy and speed of Nd:YAG laser in cutting twelve different threads were studied. Results-The guillotine cut thread faster (p < 0.05) than standard suture cutters. Blackloading thread shortened time taken to cut thread (p < 0.001) compared with free-hand cutting. Nd:YAG laser was ineffective in cutting uncolored threads and slower than mechanical cutters. Results of thread cutting in clinical studies using sewing machine (n = 77 cutting episodes in 21 patients), in-vivo experiments (n = 156), and postsurgical cases (n = 15 over 15 years) are presented. Conclusions: New thread-cutting methods are described and their efficacy demonstrated in experimental and clinical studies.
引用
收藏
页码:667 / 674
页数:8
相关论文
共 14 条
[1]   OPERATIVE ENDOSCOPY - DIATHERMY-HOOK FOR CUTTING OF NONABSORBABLE SUTURE MATERIAL [J].
CLASSEN, M ;
ROSCH, W ;
FRUHMORG.P ;
SEUBERTH, K ;
DEMLING, L .
ENDOSCOPY, 1974, 6 (01) :42-44
[2]   SUTURE-LINE ULCER AFTER GASTRIC SURGERY CAUSED BY NON-ABSORBABLE SUTURE MATERIALS [J].
GEAR, MWL ;
DOWLING, BL .
BRITISH JOURNAL OF SURGERY, 1970, 57 (05) :356-&
[3]   A NEW DETACHABLE SNARE FOR HEMOSTASIS IN THE REMOVAL OF LARGE POLYPS OR OTHER ELEVATED LESIONS [J].
HACHISU, T .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1991, 5 (02) :70-74
[4]   REFLUX CONTROL USING ENDOLUMINAL SUTURING AT GASTROSCOPY - EARLY RESULTS IN MAN [J].
KADIRKAMANATHAN, SS ;
EVANS, DF ;
GONG, F ;
HEPWORTH, CC ;
SWAIN, CP .
GASTROINTESTINAL ENDOSCOPY, 1995, 41 (04) :352-352
[5]  
Melzer A, 1994, Endosc Surg Allied Technol, V2, P78
[6]   LONG-TERM COMPARATIVE STUDY OF NONABSORBABLE SUTURES [J].
POSTLETHWAIT, RW .
ANNALS OF SURGERY, 1970, 171 (06) :892-+
[7]  
ROESCH W, 1970, ENDOSCOPY, V2, P237
[8]   APPLIANCES FOR SURGICAL ENDOSCOPY [J].
SEIFERT, E .
ENDOSCOPY, 1973, 5 (02) :96-99
[9]   SUTURE LINE ULCER AFTER GASTRIC SURGERY [J].
SMALL, WP ;
SMITH, AN ;
FALCONER, CW ;
SIRCUS, W ;
BRUCE, J .
AMERICAN JOURNAL OF SURGERY, 1968, 115 (04) :477-&
[10]   AN ENDOSCOPICALLY DELIVERABLE TISSUE-TRANSFIXING DEVICE FOR SECURING BIOSENSORS IN THE GASTROINTESTINAL-TRACT [J].
SWAIN, CP ;
BROWN, GJ ;
GONG, F ;
MILLS, TN .
GASTROINTESTINAL ENDOSCOPY, 1994, 40 (06) :730-734