Comparison of the Sideris and Amplatzer Septal Occlusion devices

被引:48
作者
Walsh, KP [1 ]
Tofeig, M [1 ]
Kitchiner, DJ [1 ]
Peart, I [1 ]
Arnold, R [1 ]
机构
[1] Alder Hey Childrens Hosp, Liverpool L12 2AP, Merseyside, England
关键词
D O I
10.1016/S0002-9149(98)01059-5
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The results of transcatheter atrial septal defect (ASD) occlusion with 2 different devices (Sideris adjustable buttoned device vs Amplatzer Septal Occluder) were compared in 2 consecutive series of patients. Comparative outcomes were assessed by whether a device was implanted or not, by complications and fluoroscopy time of implantation, and by the incidence of residual shunting on transthoracic echocardiography at follow-up. The patient and defect characteristics were similar in both groups. Twenty-eight of 33 Sideris devices and 37 of 39 Amplatzer devices were implanted, The fluoroscopy time for the Amplatzer implants was 13.4 minutes (range 8 to 41) compared with 23.7 minutes (range 11 to 60.6) for the Sideris implants (p <0.001). The complete occlusion rate for the Amplatzer device was 93% compared with 44% for the Sideris device at 1 year (p <0.001), In conclusion, the Amplatzer device produces higher occlusion rates of ASDs with shorter fluoroscopy times, (C) 1999 by Excerpta Medico, Inc.
引用
收藏
页码:933 / 936
页数:4
相关论文
共 17 条
  • [1] ARABIA FA, 1993, J THORAC CARDIOV SUR, V106, P886
  • [2] TRANSCATHETER CLOSURE OF ATRIAL SEPTAL-DEFECTS
    BABIC, UU
    GRUJICIC, S
    DJURISIC, Z
    VUCINIC, M
    [J]. LANCET, 1990, 336 (8714) : 566 - 567
  • [3] Bjornstad PG, 1997, CARDIOL YOUNG, V7, P277
  • [4] EXPERIMENTAL ATRIAL SEPTAL-DEFECT CLOSURE WITH A NEW, TRANSCATHETER, SELF-CENTERING DEVICE
    DAS, GS
    VOSS, G
    JARVIS, G
    WYCHE, K
    GUNTHER, R
    WILSON, RF
    [J]. CIRCULATION, 1993, 88 (04) : 1754 - 1764
  • [5] Transcatheter closure of atrial septal defects by the ''buttoned'' device: Complications and need for surgical revision
    Gildein, HP
    Dabritz, S
    Geibel, A
    Sarai, K
    VazquezJimenez, J
    Hugel, W
    Mocellin, R
    [J]. PEDIATRIC CARDIOLOGY, 1997, 18 (05) : 328 - 331
  • [6] Surgical repair of a large residual atrial septal defect after transcatheter closure
    Hekmat, K
    Mehlhorn, U
    deVivie, ER
    [J]. ANNALS OF THORACIC SURGERY, 1997, 63 (05) : 1456 - 1458
  • [7] Clinical impact of transcatheter closure of secundum atrial septal detects with the double umbrella device
    Justo, RN
    Nykanen, DG
    Boutin, C
    McCrindle, BW
    Freedom, RM
    Benson, LN
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 1996, 77 (10) : 889 - &
  • [8] SECUNDUM ATRIAL SEPTAL-DEFECT - NONOPERATIVE CLOSURE DURING CARDIAC-CATHETERIZATION
    KING, TD
    THOMPSON, SL
    STEINER, C
    MILLS, NL
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1976, 235 (23): : 2506 - 2509
  • [9] TRANSCATHETER CLOSURE OF ATRIAL SEPTAL-DEFECTS - EXPERIMENTAL STUDIES
    LOCK, JE
    ROME, JJ
    DAVIS, R
    VANPRAAGH, S
    PERRY, SB
    VANPRAAGH, R
    KEANE, JF
    [J]. CIRCULATION, 1989, 79 (05) : 1091 - 1099
  • [10] Closure of atrial septal defects by transcatheter devices
    Magee, AG