A direct (Non-Competitive) immunoassay for gentamicin residues with an optical biosensor

被引:24
作者
Haasnoot, W [1 ]
Verheijen, R [1 ]
机构
[1] State Inst Qual Control Agr Prod, NL-6708 PD Wageningen, Netherlands
关键词
direct immunoassay; gentamicin; optical biosensor; BIACORE; 3000;
D O I
10.1080/09540100120055701
中图分类号
O69 [应用化学];
学科分类号
081704 ;
摘要
Using sensor chip immobilized anti-gentamicin monoclonal antibodies in an optical biosensor (BIACORE 3000) resulted in the first single antibody based and label free noncompetitive immunoassay for the direct detection of residues of a low molecular weight compound in a food product. Calibration curves for gentamicin (Mw 466 Da) in buffer and milk showed 50% binding at 20 and 35 ng ml(-1), respectively, which is well below the maximum residue limit of 100 ng ml(-1).