An endoscopic force-position sensor grasper with minimum sensors

被引:25
作者
Dargahi, J
Najarian, S
机构
[1] Concordia Univ, Dept Mech & Ind Engn, CONCAVE Res Ctr, Montreal, PQ H3G 1M8, Canada
[2] Amirkabir Univ Technol, Fac Biomed Engn, Biomech Dept, Tehran, Iran
来源
CANADIAN JOURNAL OF ELECTRICAL AND COMPUTER ENGINEERING-REVUE CANADIENNE DE GENIE ELECTRIQUE ET INFORMATIQUE | 2003年 / 28卷 / 3-4期
基金
加拿大自然科学与工程研究理事会;
关键词
endoscopy; FEM; minimally invasive surgery; tactile sensor;
D O I
10.1109/CJECE.2003.1425102
中图分类号
TP3 [计算技术、计算机技术];
学科分类号
0812 ;
摘要
This study reports on the design, fabrication, testing, and mathematical modelling of a semiconductor microstrain-gauge endoscopic tactile sensor. The sensor can measure, with reasonable accuracy, the magnitude and the position of an applied load on a grasper. The designed assembly consists of two semiconductor microstrain-gauge sensors, which are positioned at the back face of a prototype endoscopic grasper. The intensity of the applied force can be seen on a light emitting diode (LED) device. Altogether, 20 different force magnitudes for seven different locations on the endoscopic grasper are tested experimentally. The magnitudes of force vary from 0.5 to 10 N with an increment of 0.5 N. The in-house electrical amplification system for the microstrain gauges is also designed, fabricated, and tested. The sensor is insulated and can operate safely in wet environments. It exhibits high force sensitivity, good linearity, and large dynamic range. To predict the behaviour of the designed system under various loading conditions, three-dimensional finite element modelling (FEW is used. There is a good correlation between the computed theoretical results for the force magnitudes together with their points of application and the experimental results. The miniaturized electronic device could be integrated with an endoscopic grasper.
引用
收藏
页码:155 / 161
页数:7
相关论文
共 35 条
[1]  
*ARTHR ASS N AM, 2003, SCOP MECH
[2]  
BARCOHEN Y, 2000, INT C SMART SYST ROB
[3]  
Bicchi A., 1996, Proceedings. 1996 IEEE International Conference on Robotics and Automation (Cat. No.96CH35857), P884, DOI 10.1109/ROBOT.1996.503884
[4]  
Brouwer I, 2001, ST HEAL T, V81, P69
[5]   FEEDBACK-REGULATION OF HAND GRASP OPENING AND CONTACT FORCE DURING STIMULATION OF PARALYZED MUSCLE [J].
CRAGO, PE ;
NAKAI, RJ ;
CHIZECK, HJ .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1991, 38 (01) :17-28
[6]   A micromachined piezoelectric tactile sensor for an endoscopic grasper - Theory, fabrication and experiments [J].
Dargahi, J ;
Parameswaran, M ;
Payandeh, S .
JOURNAL OF MICROELECTROMECHANICAL SYSTEMS, 2000, 9 (03) :329-335
[7]  
Dargahi J, 1999, ICRA '99: IEEE INTERNATIONAL CONFERENCE ON ROBOTICS AND AUTOMATION, VOLS 1-4, PROCEEDINGS, P299, DOI 10.1109/ROBOT.1999.769995
[8]   A piezoelectric tactile sensor with three sensing elements for robotic, endoscopic and prosthetic applications [J].
Dargahi, J .
SENSORS AND ACTUATORS A-PHYSICAL, 2000, 80 (01) :23-30
[9]   An endoscopic and robotic tooth-like compliance and roughness tactile sensor [J].
Dargahi, J .
JOURNAL OF MECHANICAL DESIGN, 2002, 124 (03) :576-582
[10]   Piezoelectric and pyroelectric transient signal analysis for detection of the temperature of a contact object for robotic tactile sensing [J].
Dargahi, J .
SENSORS AND ACTUATORS A-PHYSICAL, 1998, 71 (1-2) :89-97