Surface modification of polyurethane heart valves: effects on fatigue life and calcification

被引:19
作者
Bernacca, GM [1 ]
Wheatley, DJ [1 ]
机构
[1] Univ Glasgow, Royal Infirm, Dept Cardiac Surg, NHS Trust, Glasgow G31 2ER, Lanark, Scotland
关键词
polyurethane; surface modification; hydrodynamic function; durability; calcification;
D O I
10.1177/039139889802101202
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Polyurethane heart valves can be functionally durable with minimal calcification, in vitro. In vivo, these characteristics will depend on the resistance of the polyurethane to thrombogenesis and biodegradation. Surface modification may improve the polyurethane in these respects, but may adversely affect calcification and durability. This study investigates the effects of surface modifications of two polyurethane heart valves (PEU and PEUE) on their in vitro fatigue and calcification behaviour. Modifications included heparin, taurine, 3-aminopropyltriethoxysilane and polyethylene oxide (PEO). Neither hydrodynamic function nor leaflet thickness distribution was significantly altered by surface modification. PEO-modification was detrimental to valve fatigue durability and calcification. Heparin, taurine or aminosilane modifications of PEU valves increased durability. Aminosilane modification of PEUE valves increased durability compared with PEO modification. Appropriate surface modification may be useful to improve blood compatibility of implantable polyurethanes, and may also be advantageous as regards fatigue durability of flexing materials in longterm applications.
引用
收藏
页码:814 / 819
页数:6
相关论文
共 23 条
[1]  
BAMFORD C H, 1992, Clinical Materials, V10, P243, DOI 10.1016/0267-6605(92)90018-O
[2]  
Bernacca G M, 1992, J Heart Valve Dis, V1, P115
[3]   A DYNAMIC INVITRO METHOD FOR STUDYING BIOPROSTHETIC HEART-VALVE CALCIFICATION [J].
BERNACCA, GM ;
FISHER, AC ;
MACKAY, TG ;
WHEATLEY, DJ .
JOURNAL OF MATERIALS SCIENCE-MATERIALS IN MEDICINE, 1992, 3 (04) :293-298
[4]  
Bernacca GM, 1997, J BIOMED MATER RES, V34, P371
[5]   CALCIFICATION AND FATIGUE FAILURE IN A POLYURETHANE HEART-VALVE [J].
BERNACCA, GM ;
MACKAY, TG ;
WILKINSON, R ;
WHEATLEY, DJ .
BIOMATERIALS, 1995, 16 (04) :279-285
[6]  
Bernacca GM, 1996, J HEART VALVE DIS, V5, P538
[7]   Polyurethane heart valve durability: effects of leaflet thickness and material [J].
Bernacca, GM ;
Mackay, TG ;
Gulbransen, MJ ;
Donn, AW ;
Wheatley, DJ .
INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (06) :327-331
[8]  
BERNACCA GM, IN PRESS BIOMATERIAL
[9]  
CHU SH, 1984, BIOL BIOPROSTHETIC V, P805
[10]   THE ANTI-PLATELET ACTIVITY OF IMMOBILIZED PROSTACYCLIN [J].
EBERT, CD ;
LEE, ES ;
KIM, SW .
JOURNAL OF BIOMEDICAL MATERIALS RESEARCH, 1982, 16 (05) :629-638