Capsule Retentions and Incomplete Capsule Endoscopy Examinations: An Analysis of 2300 Examinations

被引:103
作者
Hoog, Charlotte M. [1 ]
Bark, Lars-Ake [2 ]
Arkani, Juan [3 ]
Gorsetman, Jacob [4 ]
Brostrom, Olle [1 ]
Sjoqvist, Urban [1 ]
机构
[1] Karolinska Inst, Dept Med, Stockholm Soder Hosp, S-11883 Stockholm, Sweden
[2] Karolinska Univ Hosp, Dept Gastroenterol & Hepatol, S-14186 Stockholm, Sweden
[3] Ersta Hosp, Dept Gastroenterol, S-11691 Stockholm, Sweden
[4] Danderyd Hosp, Dept Med, S-18288 Stockholm, Sweden
关键词
ICCE CONSENSUS; METAANALYSIS; MODALITIES; YIELD; RISK;
D O I
10.1155/2012/518718
中图分类号
R57 [消化系及腹部疾病];
学科分类号
100201 [内科学];
摘要
Aim. To evaluate capsule endoscopy in terms of incomplete examinations and capsule retentions and to find risk factors for these events. Material and Methods. This retrospective and consecutive study includes data from 2300 capsule enteroscopy examinations, performed at four different hospitals in Stockholm, Sweden from 2003 to 2009. Results. The frequency of incomplete examinations was 20%. Older age, male gender, suspected, and known Crohn's disease were risk factors for an incomplete examination. The PillCam capsule had the highest rate of completed examinations. Capsule retention occurred in 1.3% (n = 31). Risk factors for capsule retention were known Crohn's disease and suspected tumor. Complications of capsule retention were acute obstructive symptoms in six patients and one death related to complications after acute surgical capsule retrieval. Conclusion: Capsule endoscopy is considered a safe procedure, although obstructive symptoms and serious complications due to capsule retention can be found in a large series of patients.
引用
收藏
页数:7
相关论文
共 24 条
[1]
ICCE consensus for capsule retention [J].
Cave, D ;
Legnani, P ;
de Franchis, R ;
Lewis, BS .
ENDOSCOPY, 2005, 37 (10) :1065-1067
[2]
The risk of retention of the capsule endoscope in patients with known or suspected Crohn's disease [J].
Cheifetz, Adam S. ;
Kornbluth, Asher A. ;
Legnani, Peter ;
Schmelkin, Ira ;
Brown, Alphonso ;
Lichtiger, Simon ;
Lewis, Blair S. .
AMERICAN JOURNAL OF GASTROENTEROLOGY, 2006, 101 (10) :2218-2222
[3]
Capsule endoscopy retention: Is it a complication? [J].
Cheifetz, Adam S. ;
Lewis, Blair S. .
JOURNAL OF CLINICAL GASTROENTEROLOGY, 2006, 40 (08) :688-691
[4]
Can we predict spontaneous capsule passage after retention? A nationwide study to evaluate the incidence and clinical outcomes of capsule retention [J].
Cheon, J. H. ;
Kim, Y. -S. ;
Lee, I. -S. ;
Chang, D. K. ;
Ryu, J. -K. ;
Lee, K. J. ;
Moon, J. -S. ;
Park, C. H. ;
Kim, J. -O. ;
Shim, K. -N. ;
Choi, C. H. ;
Cheung, D. Y. ;
Jang, B. I. ;
Seo, G. -S. ;
Chun, H. -J. ;
Choi, M. -G. .
ENDOSCOPY, 2007, 39 (12) :1046-1052
[5]
ICCE consensus for bowel preparation and prokinetics [J].
de Franchis, R ;
Avgerinos, A ;
Barkin, J ;
Cave, D ;
Filoche, B .
ENDOSCOPY, 2005, 37 (10) :1040-1045
[6]
Agile patency system eliminates risk of capsule retention in patients with known intestinal strictures who undergo capsule endoscopy [J].
Herrerias, Juan M. ;
Leighton, Jonathan A. ;
Costamagna, Guido ;
Infantolino, Anthony ;
Eliakim, Rami ;
Fischer, Doron ;
Rubin, David T. ;
Manten, Howard D. ;
Scapa, Eitan ;
Morgan, Douglas R. ;
Bergwerk, Ari J. ;
Koslowsky, Binyamin ;
Adler, Samuel N. .
GASTROINTESTINAL ENDOSCOPY, 2008, 67 (06) :902-909
[7]
Ho Kevin K, 2007, Gastrointest Endosc Clin N Am, V17, P169, DOI 10.1016/j.giec.2006.11.001
[8]
Wireless capsule endoscopy [J].
Iddan, G ;
Meron, G ;
Glukhovsky, A ;
Swain, P .
NATURE, 2000, 405 (6785) :417-417
[9]
Diagnostic Accuracy of Capsule Endoscopy for Small Bowel Crohn's Disease Is Superior to That of MR Enterography or CT Enterography [J].
Jensen, Michael Dam ;
Nathan, Torben ;
Rafaelsen, Soren Rafael ;
Kjeldsen, Jens .
CLINICAL GASTROENTEROLOGY AND HEPATOLOGY, 2011, 9 (02) :124-129
[10]
Capsule retention: A feared complication of wireless capsule endoscopy [J].
Karagiannis, Stefanos ;
Faiss, Siegbert ;
Mavrogiannis, Christos .
SCANDINAVIAN JOURNAL OF GASTROENTEROLOGY, 2009, 44 (10) :1158-1165