Recalls and safety alerts involving pacemakers and implantable cardioverter-defibrillator generators

被引:135
作者
Maisel, WH [1 ]
Sweeney, MO [1 ]
Stevenson, WG [1 ]
Ellison, KE [1 ]
Epstein, LM [1 ]
机构
[1] Brigham & Womens Hosp, Dept Med, Div Cardiovasc, Cardiac Arrhythmia Serv, Boston, MA 02115 USA
来源
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION | 2001年 / 286卷 / 07期
关键词
D O I
10.1001/jama.286.7.793
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Context Unanticipated pacemaker and implantable cardioverter-defibrillator (ICD) generator malfunctions sometimes warrant recall by the US Food and Drug Administration (FDA). Despite increasingly frequent device implantation, pacemaker and ICD recalls and safety alerts (advisories) remain poorly characterized. Objectives To determine pacemaker and ICD generator advisory rates in the United States, to identify trends in these rates, and to examine their clinical and financial implications. Design and Setting Analysis of weekly FDA Enforcement Reports issued between January 1990 and December 2000 to identify all advisories involving pacemaker or ICD generators in the United States. Recalls and safety alerts involving lead malfunctions were not included. Main Outcome Measures Number of pacemakers and ICD generators in the United States subject to FDA recall or safety alert in 1990-2000; annual pacemaker and ICD advisory rates in the United States in 1990-2000; and estimated cost of device advisories. Results During the study period, 52 advisories (median [25th and 75th percentiles], 4 [4 and 7] per year) involving 408500 pacemakers and 114645 ICDs (523 145 total devices) were issued. Hardware malfunctions (35 advisories affecting 280641 devices) and computer errors (10 advisories affecting 216533 devices) accounted for 95 % of device recalls. Implantable cardioverter-clefibrillators were recalled more frequently than pacemakers (mean [SD], 16.4 [1.6] vs 6.7 [0.8] advisories per 100 person-years; P<.001). Between 1995 and 2000, the annual advisory rate increased for both pacemakers (P for trend <.001) and ICDs (P for trend =.02). An estimated 1.3 million device checks and analyses and 36187 device replacements resulted from the advisories and cost approximately $870 million. Conclusions Pacemaker and ICD recalls and safety alerts occur frequently, affect many patients, and appear to be increasing in number and rate. With the growing number of device implants and expanding indications for device therapy, the number of patients affected by device advisories will likely continue to increase.
引用
收藏
页码:793 / 799
页数:7
相关论文
共 55 条
  • [1] *AM COLL CARD, 2000, GUID CPT 2000
  • [2] Long-term follow-up of patients from a randomised trial of atrial versus ventricular pacing for sick-sinus syndrome
    Andersen, HR
    Nielsen, JC
    Thomsen, PEB
    Thuesen, L
    Mortensen, PT
    Vesterlund, T
    Pedersen, AK
    [J]. LANCET, 1997, 350 (9086) : 1210 - 1216
  • [3] Psychiatric morbidity and depressive symptomatology in patients with permanent pacemakers
    Aydemir, O
    Ozmen, E
    Kuey, L
    Kultur, S
    Yesil, M
    Postaci, N
    Bayata, S
    [J]. PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 1997, 20 (06): : 1628 - 1632
  • [4] *BUR DAT MAN STRAT, 2001, PAYM RAT I PROV OUTL
  • [5] A randomized study of the prevention of sudden death in patients with coronary artery disease
    Buxton, AE
    Lee, KL
    Fisher, JD
    Josephson, ME
    Prystowsky, EN
    Hafley, G
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 1999, 341 (25) : 1882 - 1890
  • [6] Runaway pacemaker in an implantable cardioverter defibrillator
    Carpenter, CM
    Galvin, J
    Guy, M
    McGovern, BA
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 1998, 9 (09) : 1008 - 1011
  • [7] Centers for Medicare and Medicaid Services (Cms), 2000, FED REGISTER, V65, P47054
  • [8] Canadian implantable defibrillator study (CIDS) - A randomized trial of the implantable cardioverter defibrillator against amiodarone
    Connolly, SJ
    Gent, M
    Roberts, RS
    Dorian, P
    Roy, D
    Sheldon, RS
    Mitchell, LB
    Green, MS
    Klein, GJ
    O'Brien, B
    [J]. CIRCULATION, 2000, 101 (11) : 1297 - 1302
  • [9] Effects of physiologic pacing versus ventricular pacing on the risk of stroke and death due to cardiovascular causes
    Connolly, SJ
    Kerr, CR
    Gent, M
    Roberts, RS
    Yusuf, S
    Gillis, AM
    Sami, MH
    Talajic, M
    Tang, ASL
    Klein, GJ
    Lau, C
    Newman, DM
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2000, 342 (19) : 1385 - 1391
  • [10] *FOOD DRUG ADM, 1997, FDA ENF REP