Young Investigator Award Presentation at the 13th Annual Meeting of the ESMRMB, September 1996, Prague - A proton-electron double-resonance imaging apparatus with simultaneous multiple electron paramagnetic resonance irradiation at 10 mT

被引:9
作者
Alecci, M
Lurie, DJ
Nicholson, I
Placidi, G
Sotgiu, A
机构
[1] UNIV ABERDEEN,DEPT BIOMED PHYS & BIOENGN,ABERDEEN AB9 2ZD,SCOTLAND
[2] UNIV AQUILA,DEPT SCI & TECNOL BIOMED,INFM,I-67100 LAQUILA,ITALY
来源
MAGNETIC RESONANCE MATERIALS IN PHYSICS BIOLOGY AND MEDICINE | 1996年 / 4卷 / 3-4期
关键词
proton-electron double-resonance imaging; dynamic nuclear polarization; electron paramagnetic resonance; nitroxide free radicals; resonators;
D O I
10.1007/BF01772006
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
The detection of free radicals in vivo is very important for the study of many physiologic and pathologic conditions. Free radicals have been implicated in a number of diseases such as ischemia, inflammation, kidney damage, and cancer. Proton-electron double-resonance imaging (PEDRI) allows the indirect detection of free radicals via the Overhauser effect. Nitroxide free radicals used for in vivo PEDRI studies present spectra with two or three lines, but most PEDRI experiments performed to date have used only single-line electron paramagnetic resonance (EPR) irradiation. There is theoretical evidence that simultaneous irradiation of multiple EPR transitions could increase the maximum achievable PEDRI enhancement. From the experimental point of view, this requires the combined use of a suitable multiple-frequency EPR source and a multiple-tuned EPR resonator. A novel radiofrequency (RE) triple-tuned loop-gap resonator for use in PEDRI has recently been developed, and dynamic nuclear polarization (DNP) data were reported. in the present study we describe a new PEDRI apparatus, equipped with a triple-tuned resonator, that is suitable for simultaneous double- or triple-EPR irradiation of nitroxide free radicals. in particular, the details of the EPR hardware used to generate the two or three EPR frequencies are given, and PEDRI images obtained with simultaneous multiple EPR irradiation are shown. Moreover, DNP experimental results showing the increase of the enhancement as a function of the EPR power for single and simultaneous double EPR irradiation are presented. The main goal of this apparatus is to improve the sensitivity and/or to reduce EPR irradiation power in a PEDRI experiment. This is likely to be particularly important in future biologic applications of PEDRI where the applied power must be optimized to reduce sample heating.
引用
收藏
页码:187 / 193
页数:7
相关论文
共 21 条
[1]  
ABRAGAM A, 1961, PRINCIPLES NUCLEAR M
[2]   SIMULTANEOUS 280-MHZ EPR IMAGING OF RAT ORGANS DURING NITROXIDE FREE-RADICAL CLEARANCE [J].
ALECCI, M ;
FERRARI, M ;
QUARESIMA, V ;
SOTGIU, A ;
URSINI, CL .
BIOPHYSICAL JOURNAL, 1994, 67 (03) :1274-1279
[3]   A novel multiple-tuned radiofrequency loop-gap resonator for use in PEDRI [J].
Alecci, M ;
Nicholson, I ;
Lurie, DJ .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1996, 110 (01) :82-86
[4]  
ALECCI M, 1996, P INT SOC MAGN RES M, P1384
[5]   INVIVO LOCALIZED ESR SPECTROSCOPY REFLECTING METABOLISM [J].
BACIC, G ;
NILGES, MJ ;
MAGIN, RL ;
WALCZAK, T ;
SWARTZ, HM .
MAGNETIC RESONANCE IN MEDICINE, 1989, 10 (02) :266-272
[6]   RADIOFREQUENCY FT EPR SPECTROSCOPY AND IMAGING [J].
BOURG, J ;
KRISHNA, MC ;
MITCHELL, JB ;
TSCHUDIN, RG ;
POHIDA, TJ ;
FRIAUF, WS ;
SMITH, PD ;
METCALFE, J ;
HARRINGTON, F ;
SUBRAMANIAN, S .
JOURNAL OF MAGNETIC RESONANCE SERIES B, 1993, 102 (01) :112-115
[7]   Measurement of nuclear spin [J].
Breit, G ;
Rabi, II .
PHYSICAL REVIEW, 1931, 38 (11) :2082-2083
[8]  
EATON GR, 1991, EPR IMAGING IN VIVO
[9]   INVIVO DETECTION OF INJECTED FREE-RADICALS BY OVERHAUSER EFFECT IMAGING [J].
GRUCKER, D .
MAGNETIC RESONANCE IN MEDICINE, 1990, 14 (01) :140-147
[10]   IMAGING RADIO-FREQUENCY ELECTRON-SPIN-RESONANCE SPECTROMETER WITH HIGH-RESOLUTION AND SENSITIVITY FOR INVIVO MEASUREMENTS [J].
HALPERN, HJ ;
SPENCER, DP ;
VANPOLEN, J ;
BOWMAN, MK ;
NELSON, AC ;
DOWEY, EM ;
TEICHER, BA .
REVIEW OF SCIENTIFIC INSTRUMENTS, 1989, 60 (06) :1040-1050