Real-time volumetric imaging of human heart without electrocardiographic gating by 256-detector row computed tomography - Initial experience

被引:34
作者
Kondo, C
Mori, S
Endo, M
Kusakabe, K
Suzuki, N
Hattori, A
Kusakabe, M
机构
[1] Tokyo Womens Med Univ, Dept Radiol, Div Nucl Med, Sch Med,Shinjuku Ku, Tokyo 1628666, Japan
[2] Natl Inst Radiol Sci, Dept Med Phys, Chiba 260, Japan
[3] Jikei Univ, Sch Med, Inst High Dimens Med Imaging, Res Ctr Med Sci, Tokyo, Japan
[4] Univ Fukui, Fac Engn, Fukui 910, Japan
关键词
computed tomography; cone beam; coronary angiography; myocardial perfusion; ventricular function;
D O I
10.1097/01.rct.0000173844.89988.37
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Objective: The feasibility of human cardiac imaging using a prototype 256-detector row cone-beam computed tomography (256CBCT) scanner without electrocardiographic gating was examined. Methods: Two healthy male volunteers were examined by contrast-enhanced 256CBCT. The number of detectors was 912 x 256, each measuring approximately 0.5 mm x 0.5 mm at the center of rotation. The craniocaudal coverage was approximately 100 mm after reconstruction by the Feldkamp-Davis-Kress algorithm. The effective time resolution was 500 milliseconds using a half-scan algorithm. Results: Serial enhancement of the left ventricular myocardium was detected. The right and left coronary arteries at proximal and distal segments were depicted without significant blurring. Although the left ventricular wall motion on cine images was not smooth over time, it was possible to measure ventricular volume and ejection fraction. Conclusions: Using the 256CBCT, it was possible to visualize the coronary arteries, myocardial perfusion, and ventricular contraction simultaneously during a single acquisition.
引用
收藏
页码:694 / 698
页数:5
相关论文
共 14 条
[1]  
Carrascosa PM, 2004, CIRCULATION, V110, P446
[2]  
Endo M, 1998, Radiat Med, V16, P7
[3]   PRACTICAL CONE-BEAM ALGORITHM [J].
FELDKAMP, LA ;
DAVIS, LC ;
KRESS, JW .
JOURNAL OF THE OPTICAL SOCIETY OF AMERICA A-OPTICS IMAGE SCIENCE AND VISION, 1984, 1 (06) :612-619
[4]   Detection of myocardial ischemia by electron beam CT - Experimental studies [J].
Knollmann, FD ;
Muschick, P ;
Krause, W ;
Hausmann, H ;
Hetzer, R ;
Felix, R .
ACTA RADIOLOGICA, 2001, 42 (04) :386-392
[5]   Image quality of noninvasive coronary angiography using multislice spiral computed tomography and electron-beam computed tomography -: Intraindividual comparison in an animal model [J].
Lembcke, A ;
Wiese, TH ;
Schnorr, J ;
Wagner, S ;
Mews, J ;
Kroencke, TJ ;
Enzweiler, CNH ;
Hamm, B ;
Taupitz, M .
INVESTIGATIVE RADIOLOGY, 2004, 39 (06) :357-364
[6]   Improved accuracy of noninvasive electron beam coronary angiography [J].
Lu, B ;
Shavelle, DM ;
Mao, SS ;
Chen, L ;
Child, J ;
Carson, S ;
Budoff, MJ .
INVESTIGATIVE RADIOLOGY, 2004, 39 (02) :73-79
[7]   Enlarged longitudinal dose profiles in cone-beam CT and the need for modified dosimetry [J].
Mori, S ;
Endo, M ;
Nishizawa, K ;
Tsunoo, T ;
Aoyama, T ;
Fujiwara, H ;
Murase, K .
MEDICAL PHYSICS, 2005, 32 (04) :1061-1069
[8]   Volumetric Cine Imaging for cardiovascular circulation using prototype 256-detector row computed tomography scanner (4-dimensional computed tomography) - A preliminary study with a porcine model [J].
Mori, S ;
Kondo, C ;
Suzuki, N ;
Yamashita, H ;
Hattori, A ;
Kusakabe, M ;
Endo, M .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 2005, 29 (01) :26-30
[9]   Physical performance evaluation of a 256-slice CT-scanner for four-dimensional imaging [J].
Mori, S ;
Endo, M ;
Tsunoo, T ;
Kandatsu, S ;
Tanada, S ;
Aradate, H ;
Saito, Y ;
Miyazaki, H ;
Satoh, K ;
Matsushita, S ;
Kusakabe, M .
MEDICAL PHYSICS, 2004, 31 (06) :1348-1356
[10]   OPTIMAL SHORT SCAN CONVOLUTION RECONSTRUCTION FOR FANBEAM CT [J].
PARKER, DL .