Fast gas chromatography combustion isotope ratio mass Spectrometry

被引:26
作者
Sacks, Gavin L. [1 ]
Zhang, Ying [1 ]
Brenna, J. Thomas [1 ]
机构
[1] Cornell Univ, Div Nutr Sci, Ithaca, NY 14853 USA
关键词
D O I
10.1021/ac0706325
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
We report here the first coupling of fast GC to IRMS, in a system capable of 250 ms peak widths (fwhm) at 1 mL/min flow rates, one-fifth as narrow as any previously reported GCC-IRMS system. We developed an optimized postcolumn interface that results in minimal peak broadening, using a programmable temperature vaporization injector in place of a rotary valve or backflush system to divert solvent, a narrow capillary combustion reactor followed by a cryogenic water trap with narrow-bore (< 0.20 mm We transfer lines, and a narrow W. open split to the IRMS directly inserted into the column effluent. Quantitative combustion was demonstrated with CH4 injections. A comparison of CO2 injections with different fwhm peak widths (250, 2500, and 7500 ms) showed similar precisions, SD(delta C-13) = 0.2-0.3 parts per thousand, for injections of > 600 pmol C on column; precision for the narrow peaks (250 ms) was considerably better for injections < 150 pmol C on column. SD(delta C-13) < 1 parts per thousand was achievable for injections of 5-15 pmol on column for 250 ms wide peaks, 10-fold better precision than 2500 ms wide peaks, and within a factor of 3 of the counting statistics Emit For a mixture of 15 fatty acid methyl esters (FAME), 1.5 nmol C of each on column yielded typical SD(delta C-13(pdb)) = 0.4 parts per thousand for fast GC and 0.5%o for conventional GC. For 14 of the 15 FAME, delta C-13 values between the two systems were within 1.5 parts per thousand and not significantly different Fast GCC-IRMS required one-third the run time (450 s vs 1400 s) to achieve comparable resolution. Mean peak widths for fast GCC-IRMS of the FAME were 720 ins, compared to 650 ms by fast GC with flame ionization detection. At a 15-fold dilution (100 pmol C on column for each FAME), fast GCC-IRMS achieved -2-fold better precision and accuracy than similar injections on conventional GCC-ERMS. Finally, a mixture of 10 steroids (similar to 7 nmol C (100 ng) each on column) was analyzed with mean precision of SD)(delta C-13) = 0.2 parts per thousand in 620 s by fast GCC-IRMS, while conventional GCC-IRMS required 1200 s and achieved poorer resolution. delta C-13 values for the two system were similar (Delta delta C-13 <=:5 2 parts per thousand for all steroids), indicating that accuracy is not compromised. In summary, fast GCC-IRMS can achieve similar precision to conventional GC with considerable time savings for standard sample sizes (> 1 nmol C) and achieves modest precision (similar to 1 parts per thousand) near the counting statistics limit on low level components.
引用
收藏
页码:6348 / 6358
页数:11
相关论文
共 36 条
[1]   Sourcing organic compounds based on natural isotopic variations measured by high precision isotope ratio mass spectrometry [J].
Asche, S ;
Michaud, AL ;
Brenna, JT .
CURRENT ORGANIC CHEMISTRY, 2003, 7 (15) :1527-1543
[2]  
Belov YP, 2005, AM LAB, V37, P42
[3]  
BRAITHWAITHE A, 1996, CHROMATOGRAPHIC METH
[4]   CONDENSED-PHASE CARBON ISOTOPIC STANDARDS FOR COMPOUND-SPECIFIC ISOTOPE ANALYSIS [J].
CAIMI, RJ ;
HOUGHTON, LA ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1994, 66 (18) :2989-2991
[5]  
Dagan S, 1996, J AM SOC MASS SPECTR, V7, P737, DOI 10.1016/1044-0305(96)80519-8
[6]  
FLENKER U, 2007, UNPUB ANAL CHEM
[7]   A NEW METHOD FOR PREDICTION OF BINARY GAS-PHASE DIFFUSION COEFFECIENTS [J].
FULLER, EN ;
SCHETTLE, PD ;
GIDDINGS, JC .
INDUSTRIAL AND ENGINEERING CHEMISTRY, 1966, 58 (05) :19-+
[8]   HIGH-SENSITIVITY TRACER DETECTION USING HIGH-PRECISION GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY AND HIGHLY ENRICHED [U-C-13]-LABELED PRECURSORS [J].
GOODMAN, KJ ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1992, 64 (10) :1088-1095
[9]   Hardware modifications to an isotope ratio mass spectrometer continuous-flow interface yielding improved signal, resolution, and maintenance [J].
Goodman, KJ .
ANALYTICAL CHEMISTRY, 1998, 70 (05) :833-837
[10]   CURVE-FITTING FOR RESTORATION OF ACCURACY FOR OVERLAPPING PEAKS IN GAS-CHROMATOGRAPHY COMBUSTION ISOTOPE RATIO MASS-SPECTROMETRY [J].
GOODMAN, KJ ;
BRENNA, JT .
ANALYTICAL CHEMISTRY, 1994, 66 (08) :1294-1301