An imaging CdZnTe detector with coplanar orthogonal anode strips

被引:2
作者
Hamel, LA [1 ]
Tousignant, O [1 ]
Couillard, M [1 ]
Courville, JF [1 ]
Jordanov, VT [1 ]
Macri, JR [1 ]
Larson, K [1 ]
Mayer, M [1 ]
McConnell, ML [1 ]
Ryan, JM [1 ]
机构
[1] Univ Montreal, Grp Rech Phys & Technol Couches Minces, GCM, Montreal, PQ H3C 3J7, Canada
来源
SEMICONDUCTORS FOR ROOM-TEMPERATURE RADIATION DETECTOR APPLICATIONS II | 1997年 / 487卷
关键词
D O I
10.1557/PROC-487-211
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
A novel electrode configuration for CZT imaging devices is presented. It is made of focusing, non-collecting anode strips, in one dimension, and collecting anode pixels, interconnected in rows, in the orthogonal dimension. The simulation of such an imaging detector is presented. First, field lines in the detector are computed that show that electrons generated in gamma-ray events are collected on the pixels. Charge signals, induced on the pixel and on the strip by drifting electrons, we calculated for several points of interaction inside the detector unit cell. These show that this new detector should retain the spectroscopic and detection efficiency advantages of single carrier (electrons) charge sensing devices such as pixel detectors or spectrometers with controlling electrodes. Furthermore, it retains the main advantage of conventional strip detectors, i.e. an N x N array of imaging pixels realized with only 2N electronic channels. An additional potential advantage is the measurement of the third coordinate, i.e. the depth of interaction in the detector.