Comparison of quadrupole and (quadrupole) ion-trap mass spectrometers for the analysis of benzodiazepines

被引:7
作者
Borrey, D [1 ]
Meyer, E [1 ]
Lambert, W [1 ]
De Leenheer, AP [1 ]
机构
[1] State Univ Ghent, Toxicol Lab, B-9000 Ghent, Belgium
关键词
mass spectrometry; benzodiazepines;
D O I
10.1016/S0021-9673(98)00450-6
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
GC-MS with on-column injection is evaluated for the analysis of benzodiazepines. A standard mixture of five benzodiazepines (ketazolam, flunitrazepam, flurazepam, alprazolam and triazolam) was used to compare the applicability of different types of mass spectrometers. Using an ion-trap mass spectrometer excessive peak tailing was observed on the late-eluting peaks. Injections on four other ion-trap systems, both with internal or external ionisation, yielded similar results. However, all peak shapes became acceptable when injections were performed on a quadrupole system. The results obtained on three other quadrupole mass spectrometers confirmed this observation. These data strongly suggest that a quadrupole instrument is more suitable than an ion-trap one for the analysis of some benzodiazepines. (C) 1998 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:125 / 131
页数:7
相关论文
共 13 条
[1]  
Altman DG., 1995, PRACTICAL STAT MED R
[2]   URINALYSIS OF ALPHA-HYDROXYALPRAZOLAM, ALPHA-HYDROXYTRIAZOLAM, AND OTHER BENZODIAZEPINE COMPOUNDS BY GC/EIMS [J].
DICKSON, PH ;
MARKUS, W ;
MCKERNAN, J ;
NIPPER, HC .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1992, 16 (02) :67-71
[3]  
DOLAN JW, 1997, LC GC EUR, V10, P788
[4]   Comparison of an ion-trap and a quadrupole mass spectrometer using diazepam as a model compound [J].
Fitzgerald, RL ;
ONeal, CL ;
Hart, BJ ;
Poklis, A ;
Herold, DA .
JOURNAL OF ANALYTICAL TOXICOLOGY, 1997, 21 (06) :445-450
[5]  
GROB K, 1991, ON COLUMN INJECTION
[6]   The quadrupole ion trap mass spectrometer - A small solution to a big challenge [J].
Jonscher, KR ;
Yates, JR .
ANALYTICAL BIOCHEMISTRY, 1997, 244 (01) :1-15
[7]   PROCEDURE FOR VERIFICATION OF FLUNITRAZEPAM AND NITRAZEPAM INTAKE BY GAS CHROMATOGRAPHIC-MASS SPECTROMETRIC ANALYSIS OF URINE [J].
LIN, Z ;
BECK, O .
JOURNAL OF PHARMACEUTICAL AND BIOMEDICAL ANALYSIS, 1995, 13 (4-5) :719-722
[8]   Simultaneous determination of midazolam and its metabolites 1-hydroxymidazolam and 4-hydroxymidazolam in human serum using gas chromatography mass spectrometry [J].
Martens, J ;
Banditt, P .
JOURNAL OF CHROMATOGRAPHY B, 1997, 692 (01) :95-100
[9]  
*MAT, 1992, MAGN GC MS SYST OP M
[10]  
MCLUCKEY SA, 1994, ANAL CHEM, V66, P689