Reconstruction from truncated projections in CT using adaptive detruncation

被引:78
作者
Sourbelle, K [1 ]
Kachelriess, M [1 ]
Kalender, WA [1 ]
机构
[1] Univ Erlangen Nurnberg, Inst Med Phys, D-91052 Erlangen, Germany
关键词
artifact correction; CT; truncated projections;
D O I
10.1007/s00330-004-2621-9
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
If the object exceeds the field of measurement (FOM) of a given CT scanner, severe artifacts may result. In this work, we propose an adaptive detruncation (ADT) method to reconstruct images from medical CT projections which are truncated in the transaxial direction. The truncated projections are extrapolated by estimating the convex hull of the patient. The ADT method allows us not only to achieve artifact-free images in the FOM but also to extend the images beyond the FOM, and can therefore be very attractive, for example, in PET/CT scanners for attenuation correction.
引用
收藏
页码:1008 / 1014
页数:7
相关论文
共 6 条
[1]  
HSIEH J, 2003, RADIOLOGY, V229, P168
[2]   Kymogram detection and kymogram-correlated image reconstruction from subsecond spiral computed tomography scans of the heart [J].
Kachelriess, M ;
Sennst, DA ;
Maxlmoser, W ;
Kalender, WA .
MEDICAL PHYSICS, 2002, 29 (07) :1489-1503
[3]   TRUNCATION ARTIFACT REDUCTION IN TRANSMISSION CT FOR IMPROVED SPECT ATTENUATION COMPENSATION [J].
KADRMAS, DJ ;
JASZCZAK, RJ ;
MCCORMICK, JW ;
COLEMAN, RE ;
LIM, CB .
PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (06) :1085-1104
[4]  
LEWITT R M, 1979, Medical Physics (Woodbury), V6, P412, DOI 10.1118/1.594519
[5]   Efficient correction for CT image artifacts caused by objects extending outside the scan field of view [J].
Ohnesorge, B ;
Flohr, T ;
Schwarz, K ;
Heiken, JP ;
Bae, KT .
MEDICAL PHYSICS, 2000, 27 (01) :39-46
[6]  
SOURBELLE K, 2003, RADIOLOGY P, V229, P692