Effect of geometrical distortion correction in MR on image registration accuracy

被引:76
作者
Maurer, CR
Aboutanos, GB
Dawant, BM
Gadamsetty, S
Margolin, RA
Maciunas, RJ
Fitzpatrick, JM
机构
[1] VANDERBILT UNIV, DEPT BIOMED ENGN, NASHVILLE, TN 37212 USA
[2] VANDERBILT UNIV, DEPT ELECTR ENGN, NASHVILLE, TN 37212 USA
[3] VANDERBILT UNIV, DEPT PSYCHIAT, NASHVILLE, TN 37212 USA
[4] VANDERBILT UNIV, DEPT RADIOL, NASHVILLE, TN 37212 USA
[5] VANDERBILT UNIV, DEPT NEUROL SURG, NASHVILLE, TN 37212 USA
关键词
image registration; magnetic resonance imaging; physics and instrumentation; image correction; techniques;
D O I
10.1097/00004728-199607000-00032
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
In this article we investigate the effect of geometrical distortion correction in MR images on the accuracy of the registration of X-ray CT and MR head images for both a fiducial marker (extrinsic point) method and a surface-matching technique. We use CT and T2-weighted MR image volumes acquired from seven patients who underwent craniotomies in a stereotactic neurosurgical clinical trial. Each patient had four external markers attached to transcutaneous posts screwed into the outer table of the skull. The MR images are corrected for static field inhomogeneity by using an image rectification technique and corrected for scale distortion (gradient magnitude uncertainty) by using an attached stereotactic frame as an object of known shape and size, We define target registration error (TRE) as the distance between corresponding marker positions after registration and transformation. The accuracy of the fiducial marker method is determined by using each combination of three markers to estimate the transformation and the remaining marker to calculate registration error. Surface-based registration is accomplished by fitting MR contours corresponding to the CSF-dura interface to CT contours derived from the inner surface of the skull. The mean point-based TRE using three noncollinear fiducials improved 34%-from 1.15 to 0.76 mm-after correcting for both static field inhomogeneity and scale distortion. The mean surface-based TRE improved 46%-from 2.20 to 1.19 mm. Correction of geometrical distortion in MR images can significantly improve the accuracy of point-based and surface-based registration of CT and MR head images. Distortion correction can be important in clinical situations such as stereotactic and functional neurosurgery where 1 to 2 mm accuracy is required.
引用
收藏
页码:666 / 679
页数:14
相关论文
共 86 条
[1]  
ABOUTANOS GB, 1993, C159303 VAND U CTR I
[2]  
ALPERT NM, 1990, J NUCL MED, V31, P1717
[3]   LEAST-SQUARES FITTING OF 2 3-D POINT SETS [J].
ARUN, KS ;
HUANG, TS ;
BLOSTEIN, SD .
IEEE TRANSACTIONS ON PATTERN ANALYSIS AND MACHINE INTELLIGENCE, 1987, 9 (05) :699-700
[5]  
BHAGWANDIEN R, 1994, THESIS UTRECHT U UTR
[6]   SUSCEPTIBILITY-LIMITED RESOLUTION IN NUCLEAR-MAGNETIC-RESONANCE MICROSCOPY [J].
CALLAGHAN, PT .
JOURNAL OF MAGNETIC RESONANCE, 1990, 87 (02) :304-318
[7]   A TECHNIQUE FOR ACCURATE MAGNETIC-RESONANCE-IMAGING IN THE PRESENCE OF FIELD INHOMOGENEITIES [J].
CHANG, H ;
FITZPATRICK, JM .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 1992, 11 (03) :319-329
[8]  
CHANG H, 1990, THESIS VANDERBILT U
[9]  
CHEN CT, 1988, INFORMATION PROCESSI, P601
[10]   TOTAL INHOMOGENEITY CORRECTION INCLUDING CHEMICAL-SHIFTS AND SUSCEPTIBILITY BY VIEW ANGLE TILTING [J].
CHO, ZH ;
KIM, DJ ;
KIM, YK .
MEDICAL PHYSICS, 1988, 15 (01) :7-11