Clinical helical tomotherapy commissioning dosimetry

被引:72
作者
Balog, J [1 ]
Olivera, G
Kapatoes, J
机构
[1] TomoTherapy Inc, Madison, WI 53717 USA
[2] Univ Wisconsin, Dept Phys Med, Madison, WI 53706 USA
关键词
tomotherapy; treatment planning commissioning; convolution/superposition; IMRT; end-of-planning;
D O I
10.1118/1.1625444
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Helical tomotherapy presented many unique dosimetric challenges and solutions during the initial commissioning process, and some of them are presented. The dose calculation algorithm is convolution/superposition based. This requires that the energy fluence spectrum and magnitude be quantified. The methodology for doing so is described. Aspects of the energy fluence characterization that are unique to tomotherapy are highlighted. Many beam characteristics can be measured automatically by an included megavoltage computed tomography imaging system. This greatly improves data collection efficiency. (C) 2003 American Association of Physicists in Medicine.
引用
收藏
页码:3097 / 3106
页数:10
相关论文
共 23 条
[1]   Benchmarking beam alignment for a clinical helical tomotherapy device [J].
Balog, J ;
Mackie, TR ;
Pearson, D ;
Hui, S ;
Paliwal, B ;
Jeraj, R .
MEDICAL PHYSICS, 2003, 30 (06) :1118-1127
[2]  
BALOG J, 2001, P BIOL PHYS BAS IMRT, P88
[3]  
BALOG J, 1998, THESIS U WISCONSIN M, pCH8
[4]   Characterization of the output for helical delivery of intensity modulated slit beams [J].
Balog, JP ;
Mackie, TR ;
Reckwerdt, P ;
Glass, M ;
Angelos, L .
MEDICAL PHYSICS, 1999, 26 (01) :55-64
[5]  
BLEIER AR, 1995, NOMOS CORP PUBLICATI
[6]  
Charland Paule M, 2003, J Appl Clin Med Phys, V4, P25, DOI 10.1120/1.1522990
[7]   Investigation of Kodak extended dose range (EDR) film for megavoltage photon beam dosimetry [J].
Chetty, IJ ;
Charland, PM .
PHYSICS IN MEDICINE AND BIOLOGY, 2002, 47 (20) :3629-3641
[8]   Dosimetry of therapeutic photon beams using an extended dose range film [J].
Esthappan, J ;
Mutic, S ;
Harms, WB ;
Dempsey, JF ;
Low, DA .
MEDICAL PHYSICS, 2002, 29 (10) :2438-2445
[9]  
JERAJ R, 2001, P BIOL PHYS BAS IMRT, P105
[10]   A feasible method for clinical delivery verification and dose reconstruction in tomotherapy [J].
Kapatoes, JM ;
Olivera, GH ;
Ruchala, KJ ;
Smilowitz, JB ;
Reckwerdt, PJ ;
Mackie, TR .
MEDICAL PHYSICS, 2001, 28 (04) :528-542