A comparison of mechanical strength of the femoral neck following locked intramedullary nailing using oblique versus transverse proximal screws

被引:5
作者
Aboulafia, AJ
Price, MM
Kennon, RE
Hutton, WC
机构
[1] Emory Univ, Sch Med, Dept Orthopaed Surg, Atlanta, GA USA
[2] Atlanta Vet Adm Hosp, Atlanta, GA USA
[3] Georgia Baptist Mem Med Ctr, Dept Orthopaed, Atlanta, GA USA
关键词
intramedullary nail; stress riser; mode of failure; femoral fixation;
D O I
10.1097/00005131-199903000-00002
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objective: The objective of the present study was to determine whether the 130-degree oblique or transverse proximal locking technique significantly enhances or diminishes the strength of the intramedullary nail-to-bone construct in a "stancelike" loading configuration. Design: Six paired fresh frozen femora were randomized to locked intramedullary nailing with either the 130-degree proximal locking technique or the transverse locking technique and tested in axial compressive loading. Torsion, shear, and impact loading were not tested. Patients/Participants: Six paired adult fresh frozen cadavers. Intervention: An unreamed eleven-millimeter by thirty-eight-centimeter intramedullary nail was placed under direct visualization in each femur. Proximal locking was randomly performed on each pair with either two transverse screws or a single oblique screw. Each femur was loaded to failure at a rate of 50.8 millimeters per minute (two inches per minute) by using an Instron tensiometer. Radiographs after nail insertion and after failure were obtained. Main Outcome Measures: Load to fracture and fracture pattern. Results: A significant difference between load to fracture was observed across femora of different donors (p < 0.001). No significant difference was observed in paired femora from the same donor when using either 130-degree or transverse proximal locking (p > 0.05). This result was based on nonanatomical axial loading that approximated loading during stance, without hip abductor loads to help stabilize the femur. Conclusion: Neither technique preferentially predisposed the construct to failure for the loading pattern tested. The mode of failure was the same in all cases, with failure beginning at the nail insertion site.
引用
收藏
页码:160 / 163
页数:4
相关论文
共 28 条
[1]
BOHLER J, 1968, CLIN ORTHOP RELAT R, P51
[2]
INTRAMEDULLARY NAILING OF FEMORAL-SHAFT FRACTURES .2. FRACTURE-HEALING WITH STATIC INTERLOCKING FIXATION [J].
BRUMBACK, RJ ;
UWAGIEERO, S ;
LAKATOS, RP ;
POKA, A ;
BATHON, GH ;
BURGESS, AR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (10) :1453-1462
[3]
INTRAMEDULLARY NAILING OF FEMORAL-SHAFT FRACTURES .3. LONG-TERM EFFECTS OF STATIC INTERLOCKING FIXATION [J].
BRUMBACK, RJ ;
ELLISON, TS ;
POKA, A ;
BATHON, GH ;
BURGESS, AR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1992, 74A (01) :106-112
[4]
INTRAMEDULLARY NAILING OF FEMORAL-SHAFT FRACTURES .1. DECISION-MAKING ERRORS WITH INTERLOCKING FIXATION [J].
BRUMBACK, RJ ;
REILLY, JP ;
POKA, A ;
LAKATOS, RP ;
BATHON, GH ;
BURGESS, AR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (10) :1441-1452
[5]
INTRAMEDULLARY NAILING OF OPEN FRACTURES OF THE FEMORAL-SHAFT [J].
BRUMBACK, RJ ;
ELLISON, PS ;
POKA, A ;
LAKATOS, R ;
BATHON, GH ;
BURGESS, AR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (09) :1324-1331
[6]
BURNSTEIN AH, 1972, J BONE JOINT SURG AM, V54, P1143
[7]
CLOSED INTRAMEDULLARY NAILING OF FEMUR [J].
CLAWSON, DK ;
SMITH, RF ;
HANSEN, ST .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1971, A 53 (04) :681-&
[8]
ERIKSSON E, 1985, CLIN ORTHOP RELAT R, P310
[9]
THE USE OF ONE COMPARED WITH 2 DISTAL SCREWS IN THE TREATMENT OF FEMORAL-SHAFT FRACTURES WITH INTERLOCKING INTRAMEDULLARY NAILING - A CLINICAL AND BIOMECHANICAL ANALYSIS [J].
HAJEK, PD ;
BICKNELL, HR ;
BRONSON, WE ;
ALBRIGHT, JA ;
SAHA, S .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1993, 75A (04) :519-525
[10]
Johnson K D, 1987, J Orthop Trauma, V1, P1, DOI 10.1097/00005131-198701010-00001