Creation of a uniform pleural defect model for the study of lung sealants

被引:14
作者
Araki, Masato
Tao, Hiroyuki
Sato, Toshihiko
Nakajima, Naoki
Sugai, Hajime
Hyon, Suong-Hyu
Nagayasu, Takeshi
Nakamura, Tatsuo
机构
[1] Kyoto Univ, Inst Frontier Med Sci, Dept Bioartifical Organs, Kyoto 6068507, Japan
[2] Kyoto Univ, Inst Frontier Med Sci, Dept med Si,ulat Engn, Kyoto 6068507, Japan
[3] Nagasaki Univ, Div Surg Oncol, Dept Translat Med Sci, Grad Sch Biomed Sci, Nagasaki 852, Japan
关键词
D O I
10.1016/j.jtcvs.2007.01.007
中图分类号
R5 [内科学];
学科分类号
1002 [临床医学]; 100201 [内科学];
摘要
Objective: Animal models are indispensable for the development of new therapeutic methods for the closure of alveolar air leakage. However, it is difficult to create a uniform pleural defect model. The purpose of this study was to establish an appropriate animal model for assessing the efficacy and histotoxicity of synthetic sealants for lung surgery. Methods: Nine beagle dogs were used to evaluate the pleural defect model in comparison with conventional resection procedures. A donut-shaped silicon ring with an inner diameter of 15 mm was placed on the pleura, and 0.1 mL of cyanoacrylate was dropped into the ring. A pleural defect was created by sliding a microtome blade just beneath the polymerized cyanoacrylate. Hemostasis was performed by pressure with a sponge. Results: Morphologically, round areas of the pleura were uniformly resected with our procedure. The resected tissue consisted of pleura and thin underlying lung parenchyma. Among the results from 3 surgeons, there were no significant differences in the mean time required for hemostasis (P = .69), the mean thickness of the resected tissue (P = .13), and the mean amount of air leakage from the resected area (P = .19). No penetration of cyanoacrylate into the lung parenchyma was evidenced by immunofluorescence microscopy. Histologically, when the pleura was resected without using cyanoacrylate, a thick fibrocellular layer extended to the lung parenchyma. Furthermore, severe fibrosis was observed when electrocautery was used for hemostasis. However, when the pleura was resected using cyanoacrylate, the normal alveolar structure was preserved. Conclusions: Our uniform pleural defect model using cyanoacrylate may be feasible for the evaluation of synthetic sealants for alveolar air leakage.
引用
收藏
页码:145 / 151
页数:7
相关论文
共 23 条
[1]
Prospective randomized study evaluating a biodegradable polymeric sealant for sealing intraoperative air leaks that occur during pulmonary resection [J].
Allen, MS ;
Wood, DE ;
Hawkinson, RW ;
Harpole, DH ;
McKenna, RJ ;
Walsh, GL ;
Vallieres, E ;
Miller, DL ;
Nichols, FC ;
Smythe, WR ;
Davis, RD .
ANNALS OF THORACIC SURGERY, 2004, 77 (05) :1792-1801
[2]
Cyanoacrylate glue in the treatment of ano-rectal fistulas [J].
Barillari, Paolo ;
Basso, Luigi ;
Larcinese, Antonella ;
Gozzo, Paolo ;
Indinnimeo, Marileda .
INTERNATIONAL JOURNAL OF COLORECTAL DISEASE, 2006, 21 (08) :791-794
[3]
The effect of autologous fibrin sealant (Vivostat®) on morbidity after pulmonary lobectomy:: a prospective randomised, blinded study [J].
Belboul, A ;
Dernevik, L ;
Aljassim, O ;
Skrbic, B ;
Radberg, G ;
Roberts, D .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2004, 26 (06) :1187-1191
[4]
Embolization of the nidus of brain arteriovenous malformations with n-butyl cyanoacrylate [J].
Debrun, GM ;
Aletich, V ;
Ausman, JI ;
Charbel, F ;
Dujovny, M .
NEUROSURGERY, 1997, 40 (01) :112-120
[5]
Experimental study on the in vivo behaviour of a new collagen glue in lung surgery [J].
Feito, BA ;
Rath, AM ;
Longchampt, E ;
Azorin, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2000, 17 (01) :8-13
[6]
Clinical experience with N-butyl cyanoacrylate tissue adhesive in fungal keratitis [J].
Garg, P ;
Gopinathan, U ;
Nutheti, R ;
Rao, GN .
CORNEA, 2003, 22 (05) :405-408
[7]
Experimental use of an albumin-glutaraldehyde tissue adhesive for sealing pulmonary parenchyma and bronchial anastomoses [J].
Herget, GW ;
Kassa, M ;
Riede, UN ;
Lu, Y ;
Brethner, L ;
Hasse, J .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2001, 19 (01) :4-9
[8]
IKADA Y, 1997, WOUND CLOSURE BIOMAT, P317
[9]
USE OF FIBRIN GLUE IN THORACIC-SURGERY [J].
JESSEN, C ;
SHARMA, P .
ANNALS OF THORACIC SURGERY, 1985, 39 (06) :521-524
[10]
The seating effect of fibrin glue against alveolar air leakage evaluated up to 48 h; comparison between different methods of application [J].
Kawamura, M ;
Gika, M ;
Izumi, Y ;
Horinouchi, H ;
Shinya, N ;
Mukai, M ;
Kobayashi, K .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 2005, 28 (01) :39-42