The precision of computer-generated surgical splints

被引:98
作者
Gateno, J
Xia, J
Teichgraeber, JF
Rosen, A
Hultgren, B
Vadnais, T
机构
[1] Univ Texas, Dent Branch, Dept Oral & Maxillofacial Surg, Houston, TX 77030 USA
[2] Univ Texas, Hlth Sci Ctr, Dept Oral & Maxillofacial Surg, Houston, TX USA
[3] Univ Texas, Hlth Sci Ctr, Cleft & Craniofacial Clin, Houston, TX USA
[4] Univ Texas, Hlth Sci Ctr, Sch Med, Dept Surg,Div Pediat Surg, Houston, TX USA
[5] Univ Texas, Hlth Sci Ctr, Dept Orthodont & Dentofacial Orthoped, Houston, TX USA
[6] Geodigm Corp, Minneapolis, MN USA
关键词
D O I
10.1016/S0278-2391(03)00240-4
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Purpose: The purpose of this study was to assess the precision of stereolithographic surgical splints generated by the authors' computer-aided design and manufacturing (CAD/CAM) technique by comparing them with the conventional acrylic splints. Materials and Methods: Seven volunteers were used. A pair of surgical splints, stereolithographic and conventional acrylic splints, was fabricated for each subject. A novel method was developed to quantify the airspace between the teeth and the splint. Conventional acrylic surgical splints served as a control group. The airspaces were recorded by impression materials and sliced cross-sectionally. Corresponding areas of the cross-sectional airspaces between stereolithographic and acrylic splints were measured and compared. Pearson's correlation coefficient and linear regression tests were performed. Results: Seven pairs of surgical splints were created. The areas of 98 pairs of cross-sectional airspaces were measured; The average difference between the conventional and the STL splints was 0.24 +/- 0.23 mm(2). The correlation coefficient (r) of the airspace areas between the stereolithographic and conventional acrylic splints was 1.00, and the regression coefficient (beta) was 1.03 (P < .01). Conclusions: The results indicated that the stereolithographic splints, generated by the authors' CAD/CAM technique, had a high degree of accuracy. The fit of the STL splints was the same as the conventional surgical splints. In the future, traditional plaster dental model surgery will be replaced by computer-assisted surgical planning. The surgical splints will be made in the computer and the treatment plan will be directly transferred to the patient. (C) 2003 American Association of Oral and Maxillofacial Surgeons.
引用
收藏
页码:814 / 817
页数:4
相关论文
共 7 条
[1]   SURGICAL-CORRECTION OF LONG FACE SYNDROME [J].
BELL, WH ;
CREEKMORE, TD ;
ALEXANDER, RG .
AMERICAN JOURNAL OF ORTHODONTICS AND DENTOFACIAL ORTHOPEDICS, 1977, 71 (01) :40-67
[2]  
BELL WH, 1992, MODERN PRACTICE ORTH, V1, P210
[3]   Individual prefabricated titanium implants in reconstructive craniofacial surgery:: Clinical and technical aspects of the first 22 cases [J].
Eufinger, H ;
Wehmöller, M .
PLASTIC AND RECONSTRUCTIVE SURGERY, 1998, 102 (02) :300-308
[4]   A new technique for the creation of a computerized composite skull model [J].
Gateno, J ;
Xia, J ;
Teichgraeber, JF ;
Rosen, A .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2003, 61 (02) :222-227
[5]   STEREOLITHOGRAPHIC MODELS FOR SURGICAL PLANNING - PRELIMINARY-REPORT [J].
STOKER, NG ;
MANKOVICH, NJ ;
VALENTINO, D .
JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 1992, 50 (05) :466-471
[6]   Computer-assisted three-dimensional surgical planning and simulation: 3D virtual osteotomy [J].
Xia, J ;
Ip, HH ;
Samman, N ;
Wang, DF ;
Kot, CSB ;
Yeung, RWK ;
Tideman, H .
INTERNATIONAL JOURNAL OF ORAL AND MAXILLOFACIAL SURGERY, 2000, 29 (01) :11-17
[7]  