Trace analysis of polychlorinated dibenzo-p-dioxins, dibenzofurans and WHO polychlorinated biphenyls in food using comprehensive two-dimensional gas chromatography with electron-capture detection

被引:61
作者
Danielsson, C [1 ]
Wiberg, K
Korytár, P
Bergek, S
Brinkman, UAT
Haglund, P
机构
[1] Umea Univ, Inst Environm Chem, Dept Chem, SE-90187 Umea, Sweden
[2] Netherlands Inst Fishery Res, NL-1970 AB Ijmuiden, Netherlands
[3] Free Univ Amsterdam, Dept Analyt Chem & Appl Spect, NL-1081 BT Amsterdam, Netherlands
关键词
comprehensive two-dimensional gas chromatography; GC x GC; mu ECD; PCB; PCDD/F; TEF; TEQ; quantification; food; Dioxins; GC-HRMS;
D O I
10.1016/j.chroma.2004.11.071
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Trace analysis of 2,3,7,8-polychlorinated dibenzo-p-dioxins (PCDDs), dibenzofurans (PCDFs) and the 12 WHO-PCBs (four non-ortho and eight mono-ortho congeners that have been assigned toxic equivalence factors, TEFs, by the World Health Organisation) was conducted by comprehensive two-dimensional gas chromatography with a micro electron-capture detector (GC x GC-mu ECD). Four food matrices (fish oil from herring, spiked cows' milk, vegetable oil and an eel extract) were analysed by two GC x GC laboratories, and four GC-HRMS laboratories generated reference values. The two GC x GC laboratories used different column combinations for separating the target analytes. For the first dimension, non-polar DB-XLB and VF-1 columns were used, and for the second dimension, an LC-50 liquid crystalline column with unique selectivity for planar compounds. The congener-specific and total toxic equivalence (TEQ) data obtained using DB-XLB x LC-50 were in good agreement with results obtained by the GC-HRMS laboratories. The WHO-PCB data obtained with the VF-1 x LC-50 combination was also good, but the PCDD/F concentrations were sometimes overestimated due to matrix interferences. GC x GC-mu ECD using DB-XLB x LC-50 seems to fulfil the European Community requirements of a screening method for PCDD/F and WHO-PCB TEQ in food. (c) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:61 / 70
页数:10
相关论文
共 27 条
[1]   Simple, non-moving modulation interface for comprehensive two-dimensional gas chromatography [J].
Beens, J ;
Adahchour, M ;
Vreuls, RJJ ;
van Altena, K ;
Brinkman, UAT .
JOURNAL OF CHROMATOGRAPHY A, 2001, 919 (01) :127-132
[2]   FRACTIONATION OF POLYCHLORINATED-BIPHENYLS, POLYCHLORINATED DIBENZO-PARA-DIOXINS, AND POLYCHLORINATED DIBENZOFURANS ON POROUS GRAPHITIC CARBON [J].
CREASER, CS ;
ALHADDAD, A .
ANALYTICAL CHEMISTRY, 1989, 61 (11) :1300-1302
[3]  
DANIELSSON C, 2003, ORGANOHALOGEN COMPD, V60, P395
[4]  
DEGEUS HJ, 1999, J HIGH RES CHROMATOG, V23, P189
[5]   A collaborative study of 209 PCB congeners and 6 Aroclors on 20 different HRGC columns .2. Semi-quantitative Aroclor congener distributions [J].
Frame, GM .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1997, 357 (06) :714-722
[6]   ISOLATION OF MONO-ORTHO AND NON-ORTHO POLYCHLORINATED-BIPHENYLS FROM BIOLOGICAL SAMPLES BY ELECTRON-DONOR ACCEPTOR HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHY USING A 2-(1-PYRENYL)ETHYLDIMETHYLSILYLATED SILICA COLUMN [J].
HAGLUND, P ;
ASPLUND, L ;
JARNBERG, U ;
JANSSON, B .
CHEMOSPHERE, 1990, 20 (7-9) :887-894
[7]   Shape selectivity: A key factor in comprehensive two-dimensional gas chromatographic analysis of toxic PCBs [J].
Haglund, P ;
Harju, M ;
Ong, R ;
Marriott, P .
JOURNAL OF MICROCOLUMN SEPARATIONS, 2001, 13 (07) :306-311
[8]   Comprehensive two-dimensional gas chromatography of the 209 polychlorinated biphenyls [J].
Harju, M ;
Danielsson, C ;
Haglund, P .
JOURNAL OF CHROMATOGRAPHY A, 2003, 1019 (1-2) :111-126
[9]   CRITICAL-REVIEW OF THE ANALYSIS OF NON-ORTHO-CHLOROBIPHENYLS AND MONO-ORTHO-CHLOROBIPHENYLS [J].
HESS, P ;
DEBOER, J ;
COFINO, WP ;
LEONARDS, PEG ;
WELLS, DE .
JOURNAL OF CHROMATOGRAPHY A, 1995, 703 (1-2) :417-465
[10]  
*HRGC HRMS, 1994, 1613 HRGCHRMS