S5FP:: Spectrally selective suppression with steady state free precession

被引:22
作者
Derbyshire, JA [1 ]
Herzka, DA [1 ]
McVeigh, ER [1 ]
机构
[1] NHLBI, Cardiac Energet Lab, NIH, DHHS, Bethesda, MD 20892 USA
关键词
MRI; steady-state; balanced SSFP; FISP; fat suppression;
D O I
10.1002/mrm.20633
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A method is presented that employs the inherent spectral selectivity of the Steady-State Free Precession (SSFP) pulse sequence to provide a spectral band of suppression. At TE = TR/2, SSFP partitions the magnetization into two phase-opposed spectral components. Z-storing one of these components simultaneously further excites the other, which is then suppressed by gradient crushing and RF spoiling. The Spectrally Selective Suppression with SSFP ((SFP)-F-5) method is shown to provide significant attenuation of fat signals, while the water signals are essentially unaffected and provide the normal SSFP contrast. Fat suppression is achieved with relatively little temporal overhead (less than 10% reduction in temporal resolution). (SFP)-F-5 was validated using simulations, phantoms, and human studies.
引用
收藏
页码:918 / 928
页数:11
相关论文
共 31 条
[1]   STEADY-STATE FREE PRECESSION IN NUCLEAR MAGNETIC RESONANCE [J].
CARR, HY .
PHYSICAL REVIEW, 1958, 112 (05) :1693-1701
[2]   ELIMINATION OF TRANSVERSE COHERENCES IN FLASH MRI [J].
CRAWLEY, AP ;
WOOD, ML ;
HENKELMAN, RM .
MAGNETIC RESONANCE IN MEDICINE, 1988, 8 (03) :248-260
[3]  
DARRASSE L, 1988, P 7 ANN M SMRM SAN F, P611
[4]  
Deimling M., 1994, PROC 2 ANN M SMR SAN, P495
[5]  
DERBYSHIRE J, 2002, P 10 ANN M ISMRM HON, P2359
[6]   Reduction of transient signal oscillations in True-FISP using a linear flip angle series magnetization preparation [J].
Deshpande, VS ;
Chung, YC ;
Zhang, Q ;
Shea, SM ;
Li, DB .
MAGNETIC RESONANCE IN MEDICINE, 2003, 49 (01) :151-157
[7]   3D magnetization-prepared true-FISP: A new technique for imaging coronary arteries [J].
Deshpande, VS ;
Shea, SM ;
Laub, G ;
Simonetti, OP ;
Finn, JP ;
Li, DB .
MAGNETIC RESONANCE IN MEDICINE, 2001, 46 (03) :494-502
[8]   Coronary artery magnetic resonance angiography [J].
Flamm, SD ;
Muthupillai, R .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2004, 19 (06) :686-709
[9]  
Haacke EM, 1999, MAGNETIC RESONANCE I
[10]  
HARDY CJ, 2002, P 10 ANN M ISMRM HON, P473