Non-expert use of the cerebral function monitor for neonatal seizure detection

被引:187
作者
Rennie, JM
Chorley, G
Boylan, GB
Pressler, R
Nguyen, Y
Hooper, R
机构
[1] Kings Coll Hosp London, Neonatol Unit, Dept Child Hlth, London SE5 9RS, England
[2] Kings Coll Hosp London, Dept Neurophysiol, London SE5 9RS, England
[3] Kings Coll Hosp London, Dept Publ Hlth Sci, London SE5 9RS, England
来源
ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION | 2004年 / 89卷 / 01期
关键词
D O I
10.1136/fn.89.1.F37
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
Background: The cerebral function monitor (CFM) is widely used to detect neonatal seizures, but there are very few studies comparing it with simultaneous electroencephalography (EEG). Objective: To determine the accuracy of non-expert use of the CFM and to assess interobserver agreement of CFM seizure detection. Patients: Babies admitted to the neonatal intensive care unit at King's College Hospital who were at high risk of seizure and had video-EEG monitoring. Methods: Video-EEG was used to detect seizures. Each baby had CFM recordings at speeds of 6, 15, and 30 cm/h during the EEG. Four neonatologists, trained in CFM seizure recognition, independently rated one hour CFM samples at three speeds from each baby. Interobserver agreement was quantified using Cohen's kappa. Results: CFM traces from 19 babies with EEG seizures and 21 babies without EEG seizures were analysed. Overall non-expert interpretation of the CFM performed poorly as a seizure detector compared with simultaneous EEG (sensitivities 38% at 6 cm/h; 54% at 15 cm/h; 55% at 30 cm/h). Although babies with seizures were more likely to be correctly classified at higher speeds (p = 0.02), babies without seizures were also more likely to be misclassified (p < 0.001). Agreement between observers was not good at any speed (kappa values from 0.01 to 0.39). The observers usually detected generalised seizures but often missed seizures that were focal, low amplitude, or lasted less than one minute. Conclusion: Approximately half of all neonatal seizures may be missed using CFM alone. Neonatal seizures need to be diagnosed, characterised, and quantified first using EEG. The CFM may then be useful for long term monitoring.
引用
收藏
页码:F37 / F40
页数:4
相关论文
共 14 条
  • [1] Assessment of neonatal encephalopathy by amplitude-integrated electroencephalography
    al Naqeeb, N
    Edwards, AD
    Cowan, FM
    Azzopardi, D
    [J]. PEDIATRICS, 1999, 103 (06) : 1263 - 1271
  • [2] ARCHBALD F, 1984, DEV MED CHILD NEUROL, V26, P162
  • [3] Boylan GB, 2002, ARCH DIS CHILD-FETAL, V86, P165
  • [4] Outcome of electroclinical, electrographic, and clinical seizures in the newborn infant
    Boylan, GB
    Pressler, RM
    Rennie, JM
    Morton, M
    Leow, PL
    Hughes, R
    Binnie, CD
    [J]. DEVELOPMENTAL MEDICINE AND CHILD NEUROLOGY, 1999, 41 (12) : 819 - 825
  • [5] THE EXACT ICTAL AND INTERICTAL DURATION OF ELECTROENCEPHALOGRAPHIC NEONATAL SEIZURES
    CLANCY, RR
    LEGIDO, A
    [J]. EPILEPSIA, 1987, 28 (05) : 537 - 541
  • [6] PREDICTIVE VALUE OF EARLY CONTINUOUS AMPLITUDE INTEGRATED EEG RECORDINGS ON OUTCOME AFTER SEVERE BIRTH ASPHYXIA IN FULL-TERM INFANTS
    HELLSTROMWESTAS, L
    ROSEN, I
    SVENNINGSEN, NW
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 1995, 72 (01): : F34 - F38
  • [7] Klebermass K, 2001, CHILD NERV SYST, V17, P544
  • [8] The clinical conundrum of neonatal seizures
    Levene, M
    [J]. ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2002, 86 (02): : 75 - 77
  • [9] DEVICE FOR CONTINUOUS MONITORING OF CEREBRAL ACTIVITY IN RESUSCITATED PATIENTS
    MAYNARD, D
    PRIOR, PF
    SCOTT, DF
    [J]. BRITISH MEDICAL JOURNAL, 1969, 4 (5682) : 545 - &
  • [10] MURDOCHEATON D, 1984, NEUROPAEDIATRICS, V25, P122