Peak capacity in gradient ultra performance liquid chromatography (UPLC)

被引:126
作者
Wren, SAC [1 ]
机构
[1] AstraZeneca, PAR&D, Analyt Dev, Macclesfield SK10 2NA, Cheshire, England
关键词
ultra performance liquid chromatography; gradient time; peak capacity;
D O I
10.1016/j.jpba.2004.12.028
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A prototype commercial instrument and 2.1 mm i.d. columns packed with 1.7 mu m porous particles have been used to measure peak capacity in ultra performance liquid chromatography (UPLC). Peak capacity was measured for a small molecule pharmaceutical as a function of gradient time, mobile phase flow rate, and column length. For very fast analysis, the highest peak capacity is obtained from a short column operating at high linear velocities. If an even higher peak capacity is required, a longer analysis time must be employed, and a point is reached where switching to a longer column becomes the best approach. (c) 2005 Elsevier B.V. All rights reserved.