The magnetic attraction of lower extremity external fixators in an MRI suite

被引:22
作者
Davison, BL
Cantu, RV
Van Woerkom, S
机构
[1] Grant Med Ctr, Columbus, OH USA
[2] Dartmouth Hitchcock Med Ctr, Lebanon, NH 03766 USA
关键词
MRI; external fixation;
D O I
10.1097/00005131-200401000-00005
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Objectives: To determine the magnetic attraction of 10 currently available lower extremity external fixators in the MRI suite and to test fixators with minimal or no magnetic attraction for the presence of temperature elevation during a 30-minute MRI scan. Design: Consecutive testing of selected tibial external fixation models in an identical manner. Setting: Level 1 trauma center. Constructs: Ten commercially available external fixators used for lower extremity injuries applied to saw bone tibia. Intervention: External fixators; suspended in testing apparatus were placed at three locations in the MRI suite for magnetic testing. Fixators with minimal attraction were scanned for 30 minutes. Main Outcome Measurements: Magnetic attraction was measured with a digital scale, and temperature measurements were obtained with a digital thermometer and thermocouple. Results: The llizarov with carbon fiber rings, Richards Hex-Fix, and Large Synthes External Fixator had <1 kg of magnetic attraction at all three locations. No component of these fixators experienced >2degrees of temperature elevation after a 30-minute MRI scan. The EBI Dynafix with Ankle Clamp, EBI Dynafix, and EBI Dynafix Hybrid along with the Hoffman II, Hoffman II Hybrid, llizarov with stainless steel rings, and Synthes Hybrid had >1 kg of magnetic attraction at all three locations. Conclusions: Many commercially available external fixators; have components and spring clamps that have significant magnetic attraction to the MRI scanner. Fixators that have <1 kg of attraction do not experience significant heating during MRL.
引用
收藏
页码:24 / 27
页数:4
相关论文
共 26 条
[1]   THE QUALITY OF MAGNETIC-RESONANCE IMAGING, AS AFFECTED BY THE COMPOSITION OF THE HALO ORTHOSIS - A BRIEF NOTE [J].
BALLOCK, RT ;
HAJEK, PC ;
BYRNE, TP ;
GARFIN, SR .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1989, 71A (03) :431-434
[2]   HEATING EFFECTS OF METALLIC IMPLANTS BY MRI EXAMINATIONS [J].
BUCHLI, R ;
BOESIGER, P ;
MEIER, D .
MAGNETIC RESONANCE IN MEDICINE, 1988, 7 (03) :255-261
[3]   NUCLEAR MAGNETIC-RESONANCE (NMR) INVIVO STUDIES - KNOWN THRESHOLDS FOR HEALTH-EFFECTS [J].
BUDINGER, TF .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1981, 5 (06) :800-811
[4]  
CANNADA LK, 1995, CLIN ORTHOP RELAT R, P206
[5]  
CLAYMAN DA, 1990, AM J NEURORADIOL, V11, P385
[6]   POTENTIAL HAZARDS IN NMR IMAGING - HEATING EFFECTS OF CHANGING MAGNETIC-FIELDS AND RF FIELDS ON SMALL METALLIC IMPLANTS [J].
DAVIS, PL ;
CROOKS, L ;
ARAKAWA, M ;
MCREE, R ;
KAUFMAN, L ;
MARGULIS, AR .
AMERICAN JOURNAL OF ROENTGENOLOGY, 1981, 137 (04) :857-860
[7]  
DISEGI JA, 1992, AO SCI SUPPLEMENT IN, P1
[8]  
*FOOD DRUG ADM, 1997, REV GUID CHECKL ORTH, P1
[9]   MRI of cervical fixation devices: Sensation of heating caused by vibration of metallic components [J].
Hartwell, RC ;
Shellock, FG .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 1997, 7 (04) :771-772
[10]   Safety of metallic implants in magnetic resonance imaging [J].
Ho, HS .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (04) :472-477