High pressures are generated at the tip of laparoscopic graspers

被引:55
作者
Cartmill, JA
Shakeshaft, AJ
Walsh, WR
Martin, CJ
机构
[1] Univ Sydney, Nepean Hosp, Dept Surg, Penrith, NSW 2751, Australia
[2] Univ New S Wales, Prince Wales Hosp, Dept Orthopaed, Sydney, NSW, Australia
来源
AUSTRALIAN AND NEW ZEALAND JOURNAL OF SURGERY | 1999年 / 69卷 / 02期
关键词
iatrogenic trauma; laparoscopic instruments;
D O I
10.1046/j.1440-1622.1999.01496.x
中图分类号
R61 [外科手术学];
学科分类号
摘要
Background: Despite its importance the interface between tissue and instrument is poorly understood. Methods: A thin film pressure transducer was deployed between the jaws of eight different laparoscopic instruments and a simulated tissue to study the forces generated at the tip of laparoscopic graspers and the effect of a change in angle of tissue tension at the interface between tissue and laparoscopic grasper. Results: Pressure was concentrated at the tip of the instrument. Localized pressures increased on average from 210 kPa to 650 kPa as the angle of grasp was increased from 0 to 135 degrees. There was a wide variation between instruments with some exhibiting increases in tip pressure of a factor of eight as the angle was increased. Conclusion: High pressures are generated at the tip of laparoscopic graspers. Pressures increase as the angle of incidence of the grasper in relation to the tissue increases. The changes in pressure with change in angle are independent of load and handle pressure.
引用
收藏
页码:127 / 130
页数:4
相关论文
共 12 条
[2]  
BOHM B, 1995, ARCH SURG-CHICAGO, V130, P415
[3]   The movement of mucosal cells of the gallbladder within the peritoneal cavity during laparoscopic cholecystectomy [J].
Doudle, M ;
King, G ;
Thomas, WM ;
Hewett, P .
SURGICAL ENDOSCOPY-ULTRASOUND AND INTERVENTIONAL TECHNIQUES, 1996, 10 (11) :1092-1094
[4]  
Fung Y. C., 1993, BIOMECHANICS MECH PR
[5]  
GORDON JE, 1981, STRUCTURES WHY THING
[6]  
HARADA S, 1994, ACTA NEUROCHIR, P449
[7]   A COMPARISON OF THE TRAUMATIC EFFECTS OF VASCULAR CLAMPS [J].
HARVEY, JG ;
GOUGH, MH .
BRITISH JOURNAL OF SURGERY, 1981, 68 (04) :267-272
[8]   Port site recurrences after laparoscopic and thoracoscopic procedures in malignancy [J].
Johnstone, PAS ;
Rohde, DC ;
Swartz, RE ;
Fetter, JE ;
Wexner, SD .
JOURNAL OF CLINICAL ONCOLOGY, 1996, 14 (06) :1950-1956
[9]  
PATKIN M, 1993, LAPAROSCOPIC ABDOMIN
[10]   SELF-RETAINING BRAIN RETRACTOR PRESSURE DURING INTRACRANIAL PROCEDURES [J].
ROSENORN, J .
ACTA NEUROCHIRURGICA, 1987, 85 (1-2) :17-22