Capillary electrophoresis of methylmercury with injection by sample stacking

被引:44
作者
CarroDiaz, AM [1 ]
LorenzoFerreira, RA [1 ]
CelaTorrijos, R [1 ]
机构
[1] UNIV SANTIAGO COMPOSTELA,FAC QUIM,DEPT QUIM ANALIT NUTR & BROMATOL,E-15706 SANTIAGO,LA CORUNA,SPAIN
关键词
capillary electrophoresis; sample stacking; injection methods; methylmercury; organometallic compounds; cysteine;
D O I
10.1016/0021-9673(95)01133-1
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
A procedure for separation and quantitation of methylmercury by capillary electrophoresis using sample stacking as the injection technique is presented. The CE conditions have been optimized in order to separate the methylmercury from the excess cysteine peak and to concentrate large volumes of sample obtaining a low detection limit. Under the proposed operational conditions, the detection limit (S/N=3) was 12 ng g(-1) and the limit of quantitation (S/N=10) was 20 ng g(-1) with a linear range of 20-100 ng g(-1) (as methylmercury in samples). The method was tested using different reference materials with a certified methylmercury content.
引用
收藏
页码:345 / 351
页数:7
相关论文
共 16 条
  • [1] APPLICATION OF SAMPLE STACKING TO GRAVITY INJECTION IN CAPILLARY ELECTROPHORESIS
    BURGI, DS
    CHIEN, RL
    [J]. JOURNAL OF MICROCOLUMN SEPARATIONS, 1991, 3 (03) : 199 - 202
  • [2] CARRO AM, 1994, AN QUAL CONTR REF MA
  • [3] FIELD AMPLIFIED SAMPLE INJECTION IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS
    CHIEN, RL
    BURGI, DS
    [J]. JOURNAL OF CHROMATOGRAPHY, 1991, 559 (1-2): : 141 - 152
  • [4] SAMPLE STACKING OF AN EXTREMELY LARGE INJECTION VOLUME IN HIGH-PERFORMANCE CAPILLARY ELECTROPHORESIS
    CHIEN, RL
    BURGI, DS
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (09) : 1046 - 1050
  • [5] ON-COLUMN SAMPLE CONCENTRATION USING FIELD AMPLIFICATION IN CZE
    CHIEN, RL
    BURGI, DS
    [J]. ANALYTICAL CHEMISTRY, 1992, 64 (08) : A489 - A496
  • [6] DETERMINATION OF CYSTEINE IN PHARMACEUTICALS VIA LIQUID-CHROMATOGRAPHY WITH POSTCOLUMN DERIVATIZATION
    JENKE, DR
    BROWN, DS
    [J]. ANALYTICAL CHEMISTRY, 1987, 59 (11) : 1509 - 1512
  • [7] FIELD-AMPLIFIED SAMPLE STACKING IN MICELLAR ELECTROKINETIC CHROMATOGRAPHY FOR ON-COLUMN SAMPLE CONCENTRATION OF NEUTRAL MOLECULES
    LIU, ZY
    SAM, P
    SIRIMANNE, SR
    MCCLURE, PC
    GRAINGER, J
    PATTERSON, DG
    [J]. JOURNAL OF CHROMATOGRAPHY A, 1994, 673 (01) : 125 - 132
  • [8] SPECIATION OF ORGANOMERCURIALS IN MARINE SAMPLES USING CAPILLARY ELECTROPHORESIS
    MEDINA, I
    RUBI, E
    MEJUTO, MC
    CELA, R
    [J]. TALANTA, 1993, 40 (11) : 1631 - 1636
  • [9] ELECTROPHORESIS IN ONE BUFFER AT 2 PH VALUES
    MICHOV, BM
    [J]. ELECTROPHORESIS, 1989, 10 (10) : 686 - 689
  • [10] CONCENTRATION DISTRIBUTIONS IN FREE ZONE ELECTROPHORESIS
    MIKKERS, FEP
    EVERAERTS, FM
    VERHEGGEN, TPEM
    [J]. JOURNAL OF CHROMATOGRAPHY, 1979, 169 (FEB): : 1 - 10