Ultrasound velocity measurement in long bones: Measurement method and simulation of ultrasound wave propagation

被引:100
作者
Lowet, G
VanderPerre, G
机构
[1] Div. of Biomech. and Eng. Design, K.U. Leuven, 3001 Heverlee
关键词
ultrasound; bone; biomechanics; fracture healing; osteoporosis;
D O I
10.1016/0021-9290(96)00054-1
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A new method for the measurement of ultrasound velocity in long bones is presented. The method can be applied in vitro as well as in vivo. It automatically corrects for the influence of soft tissue, such that the real velocity in bone is obtained. In a series of simulation experiments, hypotheses on the followed wave path were verified. A very good agreement was found between the measurement obtained in the experimental set-up and the values calculated for the hypothesised wave path in the experimental structure. These simulations revealed the feasibility of the technique to determine the velocity in a local site of the structure. Clinical applications of this technique include the monitoring of callus consolidation after fracture and the detection of bone degenerative diseases such as osteoporosis. Copyright (C) 1996 Elsevier Science Ltd.
引用
收藏
页码:1255 / 1262
页数:8
相关论文
共 13 条
[1]   MEASUREMENT OF THE VELOCITY OF ULTRASOUND IN THE HUMAN FEMUR INVIVO [J].
ANDRE, MP ;
CRAVEN, JD ;
GREENFIELD, MA .
MEDICAL PHYSICS, 1980, 7 (04) :324-330
[2]   A CONTINUOUS WAVE TECHNIQUE FOR THE MEASUREMENT OF THE ELASTIC PROPERTIES OF CORTICAL BONE [J].
ASHMAN, RB ;
COWIN, SC ;
VANBUSKIRK, WC ;
RICE, JC .
JOURNAL OF BIOMECHANICS, 1984, 17 (05) :349-361
[3]  
BROWN SA, 1976, BIOMED ENG, V11, P124
[4]  
CRAVEN DJ, 1973, INVESTIGATIVE RADIOL, V8
[5]  
GERLANC M, 1975, CLIN ORTHOP RELAT R, V111, P175
[6]   A CONTACT METHOD FOR THE ASSESSMENT OF ULTRASONIC VELOCITY AND BROAD-BAND ATTENUATION IN CORTICAL AND CANCELLOUS BONE [J].
LANGTON, CM ;
ALI, AV ;
RIGGS, CM ;
EVANS, GP ;
BONFIELD, W .
CLINICAL PHYSICS AND PHYSIOLOGICAL MEASUREMENT, 1990, 11 (03) :243-249
[7]   THE RELATION BETWEEN RESONANT FREQUENCIES AND TORSIONAL STIFFNESS OF LONG BONES INVITRO - VALIDATION OF A SIMPLE BEAM MODEL [J].
LOWET, G ;
VANAUDEKERCKE, R ;
VANDERPERRE, G ;
GEUSENS, P ;
DEQUEKER, J ;
LAMMENS, J .
JOURNAL OF BIOMECHANICS, 1993, 26 (06) :689-696
[8]   FRACTURE STIFFNESS MEASUREMENT IN THE ASSESSMENT AND MANAGEMENT OF TIBIAL FRACTURES [J].
RICHARDSON, JB ;
KENWRIGHT, J ;
CUNNINGHAM, JL .
CLINICAL BIOMECHANICS, 1992, 7 (02) :75-79
[9]  
SIEGEL IM, 1958, SURG GYNECOL OBSTET, V107, P327
[10]   NONINVASIVE DETERMINATION OF ULNAR STIFFNESS FROM MECHANICAL RESPONSE - INVIVO COMPARISON OF STIFFNESS AND BONE-MINERAL CONTENT IN HUMANS [J].
STEELE, CR ;
ZHOU, LJ ;
GUIDO, D ;
MARCUS, R ;
HEINRICHS, WL ;
CHEEMA, C .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1988, 110 (02) :87-96