Correction for partial volume effects in regional blood flow measurements adjacent to hematomas in humans with intracerebral hemorrhage: Implementation and validation

被引:34
作者
Videen, TO
Dunford-Shore, JE
Diringer, MN
Powers, WJ
机构
[1] Washington Univ, Sch Med, Dept Neurol, E Bldg Imaging Ctr, St Louis, MO 63110 USA
[2] Washington Univ, Sch Med, Dept Radiol, St Louis, MO 63110 USA
[3] Washington Univ, Sch Med, Lillian Strauss Inst Neurosci, Jewish Hosp, St Louis, MO 63110 USA
关键词
computed tomography; image correction; brain; blood flow; brain ischemia;
D O I
10.1097/00004728-199903000-00014
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Reduced blood flow measured with PET and SPECT has been reported in brain surrounding intracerebral hemorrhage (ICH). These studies have not corrected for partial volume effects from adjacent hematomas or ventricles. We have implemented a method to correct for these effects at brain boundaries (but not at internal gray/white matter boundaries), tested it with a brain phantom, and applied it to patients. Method: Using PET and X-ray CT images of a Hoffman brain phantom containing artificial hematomas, we segmented CT images to binary images representing brain tissue and convolved these to the 3D resolution of the PET. PET images were then scaled by this net tissue contribution. Corrected images were compared with other images of the same phantom without hematomas. The same correction was then applied to human images. Results: Uncorrected phantom images had artifactual reductions surrounding hematomas that were eliminated by the correction. Corrected images,of patients with acute ICH had areas of reduced flow in white and deep gray matter, whereas overlying cortical flow was preserved. Conclusion: We have validated a method to correct for artifactual reductions in blood flow measurements at brain boundaries. Using this method, we observe areas with reduced flow near acute hematomas.
引用
收藏
页码:248 / 256
页数:9
相关论文
共 34 条
[1]   A metric for testing the accuracy of cross-modality image registration: Validation and application [J].
Black, KJ ;
Videen, TO ;
Perlmutter, JS .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1996, 20 (05) :855-861
[2]   BEAM HARDENING IN X-RAY RECONSTRUCTIVE TOMOGRAPHY [J].
BROOKS, RA ;
DICHIRO, G .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (03) :390-398
[3]  
CHAWLUK JB, 1987, J NUCL MED, V28, P431
[4]   99TCM-HMPAO SPECT STUDIES IN TRAUMATIC INTRACEREBRAL HEMATOMA [J].
CHOKSEY, MS ;
COSTA, DC ;
IANNOTTI, F ;
ELL, PJ ;
CROCKARD, HA .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 1991, 54 (01) :6-11
[5]  
DEGRADO TR, 1994, J NUCL MED, V35, P1398
[6]  
DICHIRO G, 1978, J COMPUT ASSIST TOMO, V2, P65
[7]   PET STUDIES OF CORTICAL DIASCHISIS IN PATIENTS WITH MOTOR HEMI-NEGLECT [J].
FIORELLI, M ;
BLIN, J ;
BAKCHINE, S ;
LAPLANE, D ;
BARON, JC .
JOURNAL OF THE NEUROLOGICAL SCIENCES, 1991, 104 (02) :135-142
[8]  
GROTTA J, 1995, NEUROLOGY, V45, pA411
[9]   CORRECTION OF POSITRON EMISSION TOMOGRAPHY DATA FOR CEREBRAL ATROPHY [J].
HERSCOVITCH, P ;
AUCHUS, AP ;
GADO, M ;
CHI, D ;
RAICHLE, ME .
JOURNAL OF CEREBRAL BLOOD FLOW AND METABOLISM, 1986, 6 (01) :120-124
[10]  
HERSCOVITCH P, 1983, J NUCL MED, V24, P782