Clinical evaluation of a high-resolution new peripheral quantitative computerized tomography (pQCT) scanner for the bone densitometry at the lower limbs

被引:49
作者
Braun, MJ
Meta, MD
Schneider, P
Reiners, C
机构
[1] Univ Wurzburg, Clin Nucl Med, D-97080 Wurzburg, Germany
[2] Univ Buenos Aires, Fac Dent, Dept Physiol, RA-1053 Buenos Aires, DF, Argentina
关键词
D O I
10.1088/0031-9155/43/8/020
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Precision, long-term stability, linearity and accuracy of the x-ray peripheral quantitative computerized tomographic (pQCT) bone scanner XCT 3000 (Norland-Stratec Medical Sys.) were evaluated using the European Forearm Phantom (EFP). In vivo measurements were assessed using a standardized procedure at the distal femur and the distal tibia. hl the patient-scan mode, the spatial resolution of the system was 1.04 +/- 0.05 1p/mm as measured at the 10% level of the modulation transfer function (MTF). The contrast-detail diagram (CDD) yielded a minimal difference in attenuation coefficient (AC) of 0.07 cm(-1) at an object size of 0.5 mm. The effective dose for humans was calculated to be less than 1.5 mu Sv per scan. Short-term precision in vivo was expressed as root mean square standard deviation of paired measurements of 20 healthy volunteers (RMSSD = 0.5%). At the distal femur total volumetric density (ToD) and total cross-sectional area (ToA) were found to be less sensitive to positioning errors than at the distal tibia. Structural parameters like the polar cross-sectional moment of inertia (CSMIp) or the polar cross-sectional moment of resistance (CSMRp) showed a good short-term precision at the distal femur (RMSSD = 1.2 and 1.4%). The relation between the two skeletal sites with respect to CSMIp or CSMRp showed a high coefficient of determination (r(2) = 0.77 and 0.74).
引用
收藏
页码:2279 / 2294
页数:16
相关论文
共 46 条
[1]  
[Anonymous], 1990, ICRP PUBL
[2]  
Augat P, 1996, J BONE MINER RES, V11, P1356
[3]   Quantitative assessment of experimental fracture repair by peripheral computed tomography [J].
Augat, P ;
Merk, J ;
Genant, HK ;
Claes, L .
CALCIFIED TISSUE INTERNATIONAL, 1997, 60 (02) :194-199
[4]   Genetic variability in adult bone density among inbred strains of mice [J].
Beamer, WG ;
Donahue, LR ;
Rosen, CJ ;
Baylink, DJ .
BONE, 1996, 18 (05) :397-403
[5]   PREDICTING FEMORAL-NECK STRENGTH FROM BONE-MINERAL DATA - A STRUCTURAL APPROACH [J].
BECK, TJ ;
RUFF, CB ;
WARDEN, KE ;
SCOTT, WW ;
RAO, GU .
INVESTIGATIVE RADIOLOGY, 1990, 25 (01) :6-18
[6]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[7]   USE OF CONTRAST-DETAIL-DOSE EVALUATION OF IMAGE QUALITY IN A COMPUTED TOMOGRAPHIC SCANNER [J].
COHEN, G ;
DIBIANCA, FA .
JOURNAL OF COMPUTER ASSISTED TOMOGRAPHY, 1979, 3 (02) :189-195
[8]  
DUBBEL, 1997, HDB ENG CONSTRUCTION, pC12
[9]  
FERRETTI JL, 1995, BONE, V16, P119, DOI 10.1016/S8756-3282(94)00018-2
[10]  
Gasser J.A., 1995, BONE S, V17, P145