High-mass accuracy of product ions produced by SORI-CID using a dual electrospray ionization source coupled with FTICR mass spectrometry

被引:61
作者
Flora, JW [1 ]
Hannis, JC [1 ]
Muddiman, DC [1 ]
机构
[1] Virginia Commonwealth Univ, Dept Chem, Richmond, VA 23284 USA
关键词
D O I
10.1021/ac0011282
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
High-mass accuracy is demonstrated using internal calibration for product ions produced by sustained off-resonance irradiation collision-induced dissociation (SORICID) of a 15-mer oligonucleotide, 5'-(CTG)(5)-3'. Internal calibration for this tandem MS experiment was accomplished using a dual electrospray ionization (ESI) source coupled with Fourier transform ion cyclotron resonance mass spectrometry, (FTICR-MS) utilizing hexapole accumulation and gated trapping, The pulse sequence entails injection, trapping, and gas-phase isolation of the precursor ion of interest followed by the SORI-CID of this ion and, subsequently, injection and trapping of the internal mass calibrant (i.e., poly(ethylene glycol) with a 1000 Da average mass). The product ions and the poly(ethylene glycol) ions are then simultaneously excited by a broadband frequency chirp excitation waveform and detected. This technique corrects for space-charge effects on the measurement of an ion's cyclotron frequency experienced when externally calibrated data are used. While external calibration for FTICR-MS can result in mass errors of greater than 100 ppm, this internal standardization method demonstrated significantly more consistent accurate mass measurements with average mass errors ranging from -1.2 to -3.2 ppm for the 15-mer oligonucleotide used in this study. This method requires limited modifications to ESI-FTICR mass spectrometers and is applicable for both positive and negative modes of ionization as well as other sample types (e.g., pharmaceuticals, proteins, etc.).
引用
收藏
页码:1247 / 1251
页数:5
相关论文
共 51 条