Evaluation of an automated high-volume extraction method for viral nucleic acids in comparison to a manual procedure with preceding enrichment

被引:23
作者
Hourfar, MK [1 ]
Schmidt, M [1 ]
Seifried, E [1 ]
Roth, WK [1 ]
机构
[1] Univ Frankfurt, German Red Cross, Inst Transfus Med & Immunohaematol, D-60528 Frankfurt, Germany
关键词
automation; extraction; high-volume; NAT;
D O I
10.1111/j.1423-0410.2005.00649.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives Nucleic acid extraction still harbours the potential for improvements in automation and sensitivity of nucleic acid amplification technology (NAT) testing. This study evaluates the feasibility of a novel automated high-volume extraction protocol for NAT minipool testing in a blood bank setting. Materials and Method The chemagic Viral DNA/RNA Kit special for automated purification of viral nucleic acids from 9.6 ml of plasma by using the chemagic Magnetic Separation Module I was investigated. Analytical sensitivity for hepatitis C virus (HCV), human immunodeficiency virus-1 (HIV-1), hepatitis B virus (HBV), hepatitis A virus (HAV) and parvovirus B19 (B19) was compared to our present manual procedure that involves virus enrichment by centrifugation. Results Chemagic technology allows automation of the viral DNA/RNA extraction process. Viral nucleic acids were bound directly to magnetic beads from 9.6-ml minipools. By combining the automated magnetic beads-based extraction technology with our in-house TaqMan polymerase chain reaction (PCR) assays, 95% detection limits were 280 IU/ml for HCV, 4955 IU/ml for HIV-1, 249 IU/ml for HBV, 462 IU/ml for HAV and 460 IU/ml for B19, calculated for an individual donation in a pool of 96 donors. The detection limits of our present method were 460 IU/ml for HCV, 879 IU/ml for HIV-1, 90 IU/ml for HBV, 203 IU/ml for HAV and 314 IU/ml for B19. Conclusions The 95% detection limits obtained by using the chemagic method were within the regulatory requirements for blood donor screening. The sensitivities detected for HCV, HBV, HAV and B19 were found to be in a range similar to that of the manual purification method. Sensitivity for HIV-1, however, was found to be inferior for the chemagic method in this study.
引用
收藏
页码:71 / 76
页数:6
相关论文
共 9 条
[1]   Improved silica-guanidiniumthiocyanate DNA isolation procedure based on selective binding of bovine alpha-casein to silica particles [J].
Boom, R ;
Sol, C ;
Beld, M ;
Weel, J ;
Goudsmit, J ;
Wertheim-van Dillen, P .
JOURNAL OF CLINICAL MICROBIOLOGY, 1999, 37 (03) :615-619
[2]   First report of human immunodeficiency virus transmission via an RNA-screened blood donation [J].
Delwart, EL ;
Kalmin, ND ;
Jones, TS ;
Ladd, DJ ;
Foley, B ;
Tobler, LH ;
Tsui, RCP ;
Busch, MP .
VOX SANGUINIS, 2004, 86 (03) :171-177
[3]   Evaluation of a new PCR assay with competitive internal control sequence for blood donor screening [J].
Drosten, C ;
Weber, M ;
Seifried, E ;
Roth, WK .
TRANSFUSION, 2000, 40 (06) :718-724
[4]   TaqMan 5′-nuclease human immunodeficiency virus type IPCR assay with phage-packaged competitive internal control for high-throughput blood donor screening [J].
Drosten, C ;
Seifried, E ;
Roth, WK .
JOURNAL OF CLINICAL MICROBIOLOGY, 2001, 39 (12) :4302-4308
[5]   Automated screening of blood donations for hepatitis C virus RNA using the Qiagen BioRobot 9604 and the Roche COBAS HCV Amplicor assay [J].
Grant, PR ;
Sims, CM ;
Krieg-Schneider, F ;
Love, EM ;
Eglin, R ;
Tedder, RS .
VOX SANGUINIS, 2002, 82 (04) :169-176
[6]   Failure of routine HIV-1 tests in a case involving transmission with preseroconversion blood components during the infectious window period [J].
Ling, AE ;
Robbins, KE ;
Brown, TM ;
Dunmire, V ;
Thoe, SYS ;
Wong, SY ;
Leo, YS ;
Teo, D ;
Gallarda, J ;
Phelps, B ;
Chamberland, ME ;
Busch, MP ;
Folks, TM ;
Kalish, ML .
JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 2000, 284 (02) :210-214
[7]  
MULLERSCHULTE D, 2001, Patent No. 0843591
[8]   Feasibility and efficacy of routine PCR screening of blood donations for hepatitis C virus, hepatitis B virus, and HIV-1 in a blood-bank setting [J].
Roth, WK ;
Weber, M ;
Seifried, E .
LANCET, 1999, 353 (9150) :359-363
[9]   Yield of HCV and HIV-1NAT after screening of 3.6 million blood donations in central Europe [J].
Roth, WK ;
Weber, M ;
Buhr, S ;
Drosten, C ;
Weichert, W ;
Sireis, W ;
Hedges, D ;
Seifried, E .
TRANSFUSION, 2002, 42 (07) :862-868