Realization of a photometric scale based on cryogenic radiometry

被引:3
作者
Clare, JF [1 ]
Nield, KM [1 ]
机构
[1] Ind Res Ltd, Measurement Stand Lab, Lower Hutt, New Zealand
关键词
D O I
10.1088/0026-1394/35/4/6
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A standard of illuminance responsivity has been realized on a set of photopically-corrected detectors. It is based on the use of a cryogenic radiometer for the measurement of radiant power in terms of electrical standards and on a spectral responsivity scale established on silicon photodiodes in a "trap" configuration to evaluate the photopic response of the detectors. The spectral responsivity of the photometers was obtained from the trap detector by comparison using a spectrophotometer beam which underfilled both. The spatial responsivity of the photometers was found to be non-uniform by up to 1% and this is a significant source of uncertainty in the realization. The measurements of aperture area used visual identification of the edges and were corrected for a bias in the visual technique that was demonstrated by checks with a similar-sized ring gauge robust enough for contact measurement.
引用
收藏
页码:251 / 254
页数:4
相关论文
共 14 条
[1]  
BITTAR A, 1993, 91 IND RES LTD
[2]   Diffraction effects in the radiometry of coherent beams [J].
Butler, DJ ;
Kohler, R ;
Forbes, GW .
APPLIED OPTICS, 1996, 35 (13) :2162-2166
[3]  
CLARKE JF, 1990, 968 NZ I IND RES DEV
[4]   NATIONAL-INSTITUTE-OF-STANDARDS-AND-TECHNOLOGY DETECTOR-BASED PHOTOMETRIC SCALE [J].
CROMER, CL ;
EPPELDAUER, G ;
HARDIS, JE ;
LARASON, TC ;
PARR, AC .
APPLIED OPTICS, 1993, 32 (16) :2936-2948
[5]   TRAP DETECTORS AND THEIR PROPERTIES [J].
FOX, NP .
METROLOGIA, 1991, 28 (03) :197-202
[6]   NUMERICAL MODELING OF SILICON PHOTODIODES FOR HIGH-ACCURACY APPLICATIONS .3. INTERPOLATING AND EXTRAPOLATING INTERNAL QUANTUM-EFFICIENCY CALIBRATIONS [J].
GEIST, J ;
ROBINSON, AM ;
JAMES, CR .
JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1991, 96 (04) :481-492
[7]  
GIACOMO P, 1980, METROLOGIA, V16, P55
[8]   THE NPL RADIOMETRIC REALIZATION OF THE CANDELA [J].
GOODMAN, TM ;
KEY, PJ .
METROLOGIA, 1988, 25 (01) :29-40
[9]  
IKONEN K, 1995, METROLOGIA, V32, P689
[10]  
*INT COMM ILL, 1971, CIE PUBL, V15