Disposable strip potentiometric electrodes with solvent-polymeric ion-selective membranes fabricated using screen-printing technology

被引:51
作者
Koncki, R
Glab, S
Dziwulska, J
Palchetti, I
Mascini, M
机构
[1] Univ Warsaw, Dept Chem, PL-02093 Warsaw, Poland
[2] Univ Florence, Dipartimento Sanita Pubbl, Sez Chim Analit, I-50121 Florence, Italy
关键词
potentiometry; ion-selective electrodes; thick-film sensors; screen-printing technology;
D O I
10.1016/S0003-2670(98)00726-0
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A simple and reproducible method for mass production of disposable, plastic, thick-film, ion-selective electrodes using screen-printing technology is described. The properties of selective sensors for potassium, ammonium and nitrate ions obtained with this technology are shown and discussed. All the obtained strip sensors have analytical characteristics (sensitivity, selectivity, dynamic range and response time) comparable with those for conventional ion-selective electrodes. However, they practically do not require any conditioning before measurements and they respond in short time (steady-state signal is reached after a few seconds), that is important in case of the disposable sensors. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:451 / 459
页数:9
相关论文
共 22 条
[1]   DISPOSABLE SODIUM ELECTRODES [J].
BORCHARDT, M ;
DIEKMANN, C ;
DUMSCHAT, C ;
CAMMANN, K ;
KNOLL, M .
TALANTA, 1994, 41 (06) :1025-1028
[2]   ELECTROCHEMICAL PERFORMANCE, BIOCOMPATIBILITY, AND ADHESION OF NEW POLYMER MATRICES FOR SOLID-STATE ION SENSORS [J].
CHA, GS ;
LIU, D ;
MEYERHOFF, ME ;
CANTOR, HC ;
MIDGLEY, AR ;
GOLDBERG, HD ;
BROWN, RB .
ANALYTICAL CHEMISTRY, 1991, 63 (17) :1666-1672
[3]   ELECTROANALYTICAL AND BIOCOMPATIBILITY STUDIES ON CARBOXYLATED POLY(VINYL CHLORIDE) MEMBRANES FOR MICROFABRICATED ARRAY SENSORS [J].
COSOFRET, VV ;
ERDOSY, M ;
BUCK, RP ;
KAO, WJ ;
ANDERSON, JM ;
LINDNER, E ;
NEUMAN, MR .
ANALYST, 1994, 119 (11) :2283-2292
[4]   DOUBLE MATRIX MEMBRANES FOR POTENTIOMETRIC CATION-SELECTIVE DISPOSABLE SENSORS [J].
DUMSCHAT, C ;
BORCHARDT, M ;
DIEKMANN, C ;
HEPKE, J ;
CAMMANN, K ;
KNOLL, M .
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY, 1994, 348 (8-9) :553-555
[5]   POTENTIOMETRIC BIOSENSOR IN DOUBLE MATRIX MEMBRANE TECHNOLOGY [J].
EGGENSTEIN, C ;
BORCHARDT, M ;
DUMSCHAT, C ;
GRUNDIG, B ;
CAMMANN, K ;
SPENER, F ;
KNOLL, M .
BIOSENSORS & BIOELECTRONICS, 1995, 10 (6-7) :595-600
[6]   COMPARISON OF POTENTIOMETRIC ENZYME SENSORS FOR UREA AND PENICILLIN-G - DIFFERENTIAL THICK-FILM SENSORS VERSUS CLASSICAL ELECTRODES [J].
EPPELSHEIM, C ;
AUBECK, R ;
HAMPP, N .
JOURNAL OF MEMBRANE SCIENCE, 1995, 100 (02) :131-137
[7]   POTENTIOMETRIC THICK-FILM SENSOR FOR THE DETERMINATION OF THE NEUROTRANSMITTER ACETYLCHOLINE [J].
EPPELSHEIM, C ;
HAMPP, N .
ANALYST, 1994, 119 (10) :2167-2171
[8]   CHEMICAL SENSORS, BIOSENSORS AND THICK-FILM TECHNOLOGY [J].
GALANVIDAL, CA ;
MUNOZ, J ;
DOMINGUEZ, C ;
ALEGRET, S .
TRAC-TRENDS IN ANALYTICAL CHEMISTRY, 1995, 14 (05) :225-231
[9]   SCREEN PRINTING - A TECHNOLOGY FOR THE BATCH FABRICATION OF INTEGRATED CHEMICAL-SENSOR ARRAYS [J].
GOLDBERG, HD ;
BROWN, RB ;
LIU, DP ;
MEYERHOFF, ME .
SENSORS AND ACTUATORS B-CHEMICAL, 1994, 21 (03) :171-183
[10]   DESIGN AND APPLICATION OF THICK-FILM MULTISENSORS [J].
HAMPP, N ;
EPPELSHEIM, C ;
POPP, J ;
BISENBERGER, M ;
BRAUCHLE, C .
SENSORS AND ACTUATORS A-PHYSICAL, 1992, 31 (1-3) :144-148